The Propaq LT Series monitors. Model numbers 802LTAN, 802LT0N, and 802LTRN. These devices are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO2. The most likely locations for patients to be monitored by these devices are hospital general medical-surgical, telemetry, and intermediate care floors, hospital emergency departments, transport, emergency medical services, and other healthcare applications. intended to be used by clinicians and medically qualified personnel for single or multiparameter vital signs monitoring of ambulatory and nonambulatory patients, including neonate, pediatric and adult patients. These devices are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO2. The most likely locations for patients to be monitored by these devices are hospital general medicalsurgical, telemetry, and intermediate care floors, hospital emergency departments, transport, emergency medical services, and other healthcare applications. The monitors can be used as stand-alone devices or as devices networked to an Acuity Central Station through wireless communication over a Welch Allyn FlexNet network.
Publicado el 15 de enero de 2014 · Inicio del retiro: 21 de noviembre de 2013
Resumen
FDA registra el retiro Z-0633-2014 de dispositivos médicos de Clase II, publicado el 15 de enero de 2014 e iniciado el 21 de noviembre de 2013. Empresa responsable: Welch Allyn Protocol, Inc; Estado: Finalizado; Cantidad reportada: 6003 units total (1715 units in the US and 4288 outside the US); Lote: SERIAL NUMBERS OF DEVICES DISTRIBUTED IN THE US: KA008788, KA009581, KA009685, KA009776, KA009917, KA010079, KA010193, KA010996, KA011141, KA011192, KA011227, KA011228, KA011231, KA011249, KA011303, KA011307, KA011311, KA011314, KA011327, KA011355, KA011356, KA011609, KA011610, KA011611, KA011667, KA011671, KA011672, KA011673, KA011674, KA011675, KA011676, KA011677, KA011678, KA011682, KA011731, KA011758, KA020006, KA020007, KA020008, KA020009, KA020010, KA020011, KA020012, KA020013, KA020014, KA020015, KA020016, KA020018, KA020023, KA020088, KA020091, KA020092, KA020093, KA020116, KA020135, KA020140, KA020149, KA020150, KA020151, KA020192, KA020200, KA020206, KA020207, KA020208, KA020209, KA020210, KA020211, KA020212, KA020213, KA020214, KA020215, KA020262, KA020264, KA020265, KA020266, KA020269, KA020270, KA020273, KA020295, KA020299, KA020304, KA020310, KA020311, KA020313, KA020314, KA020321, KA020381, KA020385, KA020386, KA020387, KA020388, KA020389, KA020390, KA020391, KA020392, KA020393, KA020394, KA020405, KA020406, KA020407, KA020408, KA020409, KA020410, KA020411, KA020412, KA020413, KA020414, KA020417, KA020422, KA020423, KA020424, KA020425, KA020426, KA020427, KA020429, KA020430, KA020432, KA020433, KA020435, KA020436, KA020437, KA020438, KA020439, KA020442, KA020443, KA020444, KA020445, KA020446, KA020447, KA020448, KA020449, KA020450, KA020451, KA020452, KA020453, KA020454, KA020455, KA020456, KA020457, KA020458, KA020459, KA020460, KA020461, KA020462, KA020463, KA020464, KA020465, KA020466, KA020467, KA020468, KA020469, KA020470, KA020471, KA020472, KA020473, KA020474, KA020475, KA020477, KA020478, KA020479, KA020480, KA020481, KA020482, KA020483, KA020526, KA020529, KA020532, KA020533, KA020534, KA020535, KA020536, KA020537, KA020538, KA020540, KA020541, KA020542, KA020543, KA020545, KA020546, KA020547, KA020548, KA020549, KA020550, KA020551, KA020552, KA020553, KA020554, KA020555, KA020556, KA020557, KA020558, KA020559, KA020560, KA020561, KA020562, KA020563, KA020564, KA020575, KA020576, KA020577, KA020578, KA020581, KA020582, KA020595, KA020596, KA020597, KA020606, KA020611, KA020613, KA020617, KA020631, KA020632, KA020633, KA020641, KA020646, KA020649, KA020650, KA020655, KA020659, KA020666, KA020685, KA020691, KA020692, KA020693, KA020726, KA020729, KA020742, KA020743, KA020746, KA020748, KA020853, KA020908, KA020909, KA020916, KA020917, KA021019, KA021028, KA021043, KA021048, KA021061, KA021062, KA021063, KA021064, KA021065, KA021086, KA021126, KA021151, KA021155, KA021158, KA021159, KA021170, KA021181, KA021182, KA021184, KA021188, KA021190, KA021202, KA021221, KA021277, KA021278, KA021279, KA021280, KA021286, KA021294, KA021298, KA021299, KA021311, KA021312, KA021313, KA021318, KA021319, KA021322, KA021323, KA021327, KA021330, KA021349, KA021365, KA021366, KA021367, KA021368, KA021369, KA021370, KA021371, KA021392, KA021393, KA021394, KA021395, KA021396, KA021397, KA021398, KA021399, KA021401, KA021402, KA021403, KA021406, KA021407, KA021408, KA021410, KA021413, KA021415, KA021417, KA021419, KA021426, KA021427, KA021429, KA021462, KA021465, KA021466, KA021469, KA021486, KA021492, KA021495, KA021503, KA021511, KA021516, KA021522, KA021524, KA021529, KA021531, KA021534, KA021545, KA021546, KA021547, KA021549, KA021552, KA021553, KA021556, KA021557, KA021558, KA021559, KA021560, KA021562, KA021563, KA021564, KA021565, KA021566, KA021567, KA021568, KA021569, KA021570, KA021572, KA021576, KA021587, KA021588, KA021589, KA021590, KA021591, KA021592, KA021598, KA021606, KA021609, KA021611, KA021613, KA021614, KA021615, KA021616, KA021623, KA021624, KA021626, KA021630, KA021632, KA021635, KA021670, KA021679, KA021702, KA021704, KA021705, KA021706, KA021707, KA021708, KA021709, KA021711, KA021713, KA021714, KA021715, KA021716, KA021722, KA021723, KA021741, KA021742, KA021745, KA021746, KA021748, KA021749, KA021750, KA021765, KA021766, KA021786, KA021787, KA021788, KA021790, KA021799, KA021803, KA021804, KA021805. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Welch Allyn will update Propaq 802 Series Vital Signs Monitor, models are 802LTAN, 802LTON and 802LTR with software because the fault conditions relating to hardware or software failure or electrical noise caused by proximal defibrillation activity could, in rare instances, potentially interrupt device operations.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Welch Allyn Protocol, Inc
- Distribuidor
- Worldwide Distribution - USA (nationwide) and Internationally to United Arab Emirates, Argentina, Australia, Belgium, Brazil, Canada, Switzerland, Chile, China, Columbia, Costa Rica, Czech Republic, Germany, Denmark, Ecuador, Spain, France, United Kingdom, Greece, Hong Kong,Taiwan, Ireland, Italy, Japan, Korea, Republic of, Kuwait, Lithuania, Luxemborg, Mexico, Malaysia, Mozambique, Namibia, Netherlands, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Sierra Leone, El Salvador, Thailand, Turkey, Uganda, Vietnam, Kosovo, and South Africa.
- Lote
- SERIAL NUMBERS OF DEVICES DISTRIBUTED IN THE US: KA008788, KA009581, KA009685, KA009776, KA009917, KA010079, KA010193, KA010996, KA011141, KA011192, KA011227, KA011228, KA011231, KA011249, KA011303, KA011307, KA011311, KA011314, KA011327, KA011355, KA011356, KA011609, KA011610, KA011611, KA011667, KA011671, KA011672, KA011673, KA011674, KA011675, KA011676, KA011677, KA011678, KA011682, KA011731, KA011758, KA020006, KA020007, KA020008, KA020009, KA020010, KA020011, KA020012, KA020013, KA020014, KA020015, KA020016, KA020018, KA020023, KA020088, KA020091, KA020092, KA020093, KA020116, KA020135, KA020140, KA020149, KA020150, KA020151, KA020192, KA020200, KA020206, KA020207, KA020208, KA020209, KA020210, KA020211, KA020212, KA020213, KA020214, KA020215, KA020262, KA020264, KA020265, KA020266, KA020269, KA020270, KA020273, KA020295, KA020299, KA020304, KA020310, KA020311, KA020313, KA020314, KA020321, KA020381, KA020385, KA020386, KA020387, KA020388, KA020389, KA020390, KA020391, KA020392, KA020393, KA020394, KA020405, KA020406, KA020407, KA020408, KA020409, KA020410, KA020411, KA020412, KA020413, KA020414, KA020417, KA020422, KA020423, KA020424, KA020425, KA020426, KA020427, KA020429, KA020430, KA020432, KA020433, KA020435, KA020436, KA020437, KA020438, KA020439, KA020442, KA020443, KA020444, KA020445, KA020446, KA020447, KA020448, KA020449, KA020450, KA020451, KA020452, KA020453, KA020454, KA020455, KA020456, KA020457, KA020458, KA020459, KA020460, KA020461, KA020462, KA020463, KA020464, KA020465, KA020466, KA020467, KA020468, KA020469, KA020470, KA020471, KA020472, KA020473, KA020474, KA020475, KA020477, KA020478, KA020479, KA020480, KA020481, KA020482, KA020483, KA020526, KA020529, KA020532, KA020533, KA020534, KA020535, KA020536, KA020537, KA020538, KA020540, KA020541, KA020542, KA020543, KA020545, KA020546, KA020547, KA020548, KA020549, KA020550, KA020551, KA020552, KA020553, KA020554, KA020555, KA020556, KA020557, KA020558, KA020559, KA020560, KA020561, KA020562, KA020563, KA020564, KA020575, KA020576, KA020577, KA020578, KA020581, KA020582, KA020595, KA020596, KA020597, KA020606, KA020611, KA020613, KA020617, KA020631, KA020632, KA020633, KA020641, KA020646, KA020649, KA020650, KA020655, KA020659, KA020666, KA020685, KA020691, KA020692, KA020693, KA020726, KA020729, KA020742, KA020743, KA020746, KA020748, KA020853, KA020908, KA020909, KA020916, KA020917, KA021019, KA021028, KA021043, KA021048, KA021061, KA021062, KA021063, KA021064, KA021065, KA021086, KA021126, KA021151, KA021155, KA021158, KA021159, KA021170, KA021181, KA021182, KA021184, KA021188, KA021190, KA021202, KA021221, KA021277, KA021278, KA021279, KA021280, KA021286, KA021294, KA021298, KA021299, KA021311, KA021312, KA021313, KA021318, KA021319, KA021322, KA021323, KA021327, KA021330, KA021349, KA021365, KA021366, KA021367, KA021368, KA021369, KA021370, KA021371, KA021392, KA021393, KA021394, KA021395, KA021396, KA021397, KA021398, KA021399, KA021401, KA021402, KA021403, KA021406, KA021407, KA021408, KA021410, KA021413, KA021415, KA021417, KA021419, KA021426, KA021427, KA021429, KA021462, KA021465, KA021466, KA021469, KA021486, KA021492, KA021495, KA021503, KA021511, KA021516, KA021522, KA021524, KA021529, KA021531, KA021534, KA021545, KA021546, KA021547, KA021549, KA021552, KA021553, KA021556, KA021557, KA021558, KA021559, KA021560, KA021562, KA021563, KA021564, KA021565, KA021566, KA021567, KA021568, KA021569, KA021570, KA021572, KA021576, KA021587, KA021588, KA021589, KA021590, KA021591, KA021592, KA021598, KA021606, KA021609, KA021611, KA021613, KA021614, KA021615, KA021616, KA021623, KA021624, KA021626, KA021630, KA021632, KA021635, KA021670, KA021679, KA021702, KA021704, KA021705, KA021706, KA021707, KA021708, KA021709, KA021711, KA021713, KA021714, KA021715, KA021716, KA021722, KA021723, KA021741, KA021742, KA021745, KA021746, KA021748, KA021749, KA021750, KA021765, KA021766, KA021786, KA021787, KA021788, KA021790, KA021799, KA021803, KA021804, KA021805
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAAnspach B-Blue-S 11.25 cm attachment and Anspach B-Green Standard Craniotome. Cutting and shaping bone.15 de enero de 2014
- FDATracer IV wheelchairs, Heavy duty manual wheel chair, 24" seat width, 450 lbs. weight capacity. Distributed by15 de enero de 2014
- FDAAnspach BLACKMAX-N (Pneumatic motor) and Autolube III (Foot Control). Cutting and shaping bone.15 de enero de 2014
- FDASandel Healthcare Safety Devices Weighted Safety Scalpel #3 Disposable Safety Handle with #10 Blade15 de enero de 2014
- FDAStryker Color Cuff, REF 5921-034-135 Disposable Tourniquet Cuffs 34 in x 4 in (86 cm x 10 cm) Single Bladder,15 de enero de 2014
- FDASandel Healthcare Safety Devices Weighted Safety Scalpel #3 Disposable Safety Handle with #11 Blade15 de enero de 2014
Preguntas sobre este retiro
¿The Propaq LT Series monitors. Model numbers 802LTAN, 802LT0N, and 802LTRN. These devices are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO2. The most likely locations for patients to be monitored by these devices are hospital general medical-surgical, telemetry, and intermediate care floors, hospital emergency departments, transport, emergency medical services, and other healthcare applications. intended to be used by clinicians and medically qualified personnel for single or multiparameter vital signs monitoring of ambulatory and nonambulatory patients, including neonate, pediatric and adult patients. These devices are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO2. The most likely locations for patients to be monitored by these devices are hospital general medicalsurgical, telemetry, and intermediate care floors, hospital emergency departments, transport, emergency medical services, and other healthcare applications. The monitors can be used as stand-alone devices or as devices networked to an Acuity Central Station through wireless communication over a Welch Allyn FlexNet network. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0633-2014 el 15 de enero de 2014. Peligro reportado: Welch Allyn will update Propaq 802 Series Vital Signs Monitor, models are 802LTAN, 802LTON and 802LTR with software because the fault conditions relating to hardware or software failure or electrical noise caused by proximal defibrillation activity could, in rare instances, potentially interrupt device operations.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter