Incisive CT, model 728143 & 728144 running Software Version 5.0.0.
Publicado el 17 de enero de 2024 · Inicio del retiro: 23 de diciembre de 2022
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Multiple software issues which have the potential to lead to misdiagnosis due to image artifacts or incorrect image orientation labels, or the need for a CT rescan. These include: 1. Artifacts may occur on perfusion images, 2. Orientation label reversed in 2D images, 3. Auto voice does not work during the cardiac scan, 4. Auto voice and operators voice are too quiet, patient cant hear, 5. Not all of the images are reconstructed in some cases, 6. No image displayed on the console monitor after enabling CCT Single Scan, 7. Ring artifact appears in 32x0.625 collimation of bone scan when soft tissue view is selected, 8. A design defect where the error couch motion signal of couch firmware causes couch to identify current position as target position in next protocol, and 9. calcium scoring study split into separate series after loading to ISP.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Philips Healthcare (Suzhou) Co., Ltd.
- Distribuidor
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, ME, MI, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, PA, RI, SC, TN, TX, UT, VT, WI, & Puerto Rico and the countries of Albania, Algeria, Argentina, Australia, Austria, Bhutan, Bolivia, Bosnia-Herz, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Ethiopia, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Laos, Latvia, Lebanon, Lithuania, Malaysia, Mali, Mauritius, Moldova, Morocco, Netherlands, New Zealand, Norway, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Rwanda, Saint Kitts, Saint Lucia, Saudi Arabia, Senegal, Serbia, Sint Maarten, Slovakia, South Africa, South Korea, Spain, St. Pier Miquel, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, United Arab Emirates, Uzbekistan, Venezuela, & Vietnam.
- Lote
- UDI-DI: 00884838085015 (Model 728143) UDI-DI: 00884838105508 (Model 728144)
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- China