PKG, MIXTER FORCEPS, P/N 0250080285. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Publicado el 24 de diciembre de 2014 · Inicio del retiro: 17 de noviembre de 2014
Resumen
FDA registra el retiro Z-0655-2015 de dispositivos médicos de Clase II, publicado el 24 de diciembre de 2014 e iniciado el 17 de noviembre de 2014. Empresa responsable: Stryker Endoscopy; Estado: Finalizado; Cantidad reportada: 180,573 units; Lote: 41350 44269 442637 647415 743732 845623 846466 846767 847691 940788 941241 941606 942276 942920 943500 943533 945623 1040414 1041622 1042249 1042620 1042673 1042674 1042675 1046593 1047021 1140650 1142620 1143129 1144315 1144939 1145976 1146830 1240976 1242463 1244169 1246128 1246649 1341525 1341993 1342760 1343129 1346546 1346648 1440442 1442804. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Stryker Endoscopy
- Distribuidor
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Lote
- 41350 44269 442637 647415 743732 845623 846466 846767 847691 940788 941241 941606 942276 942920 943500 943533 945623 1040414 1041622 1042249 1042620 1042673 1042674 1042675 1046593 1047021 1140650 1142620 1143129 1144315 1144939 1145976 1146830 1240976 1242463 1244169 1246128 1246649 1341525 1341993 1342760 1343129 1346546 1346648 1440442 1442804
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAPKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080123 Laparoscopic Manual Instruments are intended for24 de diciembre de 2014
- FDAPKG, ENDO METZENBAUM SCISSORS, CURVED, P/N 0250080267. Laparoscopic Manual Instruments are intended for24 de diciembre de 2014
- FDAPKG, ALLIGATOR FORCEPS, P/N 0250080316. Laparoscopic Manual Instruments are intended for cutting, holding,24 de diciembre de 2014
- FDAPKG, 5MM X 33CM, INSERT, FENESTRATED GRASPER, P/N 0250080755 Laparoscopic Manual Instruments are intended for24 de diciembre de 2014
- FDAPKG, GRASPER FORCEPS, TISSUE, P/N 0250080284. Laparoscopic Manual Instruments are intended for cutting,24 de diciembre de 2014
- FDAPKG, GRASPER FORCEPS, SERRATED, P/N 0250080286. Laparoscopic Manual Instruments are intended for cutting,24 de diciembre de 2014
Preguntas sobre este retiro
¿PKG, MIXTER FORCEPS, P/N 0250080285. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0655-2015 el 24 de diciembre de 2014. Peligro reportado: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter