Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0665-2015

PKG, CANNULA, 11MM, AUTOMATIC VALVE STOP COCK, W/O TROCAR, P/N 0250080174. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Publicado el 24 de diciembre de 2014 · Inicio del retiro: 17 de noviembre de 2014

Resumen

FDA registra el retiro Z-0665-2015 de dispositivos médicos de Clase II, publicado el 24 de diciembre de 2014 e iniciado el 17 de noviembre de 2014. Empresa responsable: Stryker Endoscopy; Estado: Finalizado; Cantidad reportada: 180,573 units total; Lote: 34084 36925 38749 38885 38955 39654 44365 44521 44924 44925 45030 45806 46020 46328 46472 47128 47157 47614 47955 48064 48403 48404 48645 48699 48768 49015 49479 49832 50044 50237 50859 51025 51179 51390 51612 51910 52116 52651 52998 53778 57318 57476 58259 59578 59722 59739 61128 62393 63206 63761 64136 65374 65921 67020 68011 68985 70523 74752 75614 75641 76643 76818 76961 78384 79424 80272 81574 83533 85051 85137 85913 86043 86264 86431 89121. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Stryker Endoscopy
Distribuidor
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Lote
34084 36925 38749 38885 38955 39654 44365 44521 44924 44925 45030 45806 46020 46328 46472 47128 47157 47614 47955 48064 48403 48404 48645 48699 48768 49015 49479 49832 50044 50237 50859 51025 51179 51390 51612 51910 52116 52651 52998 53778 57318 57476 58259 59578 59722 59739 61128 62393 63206 63761 64136 65374 65921 67020 68011 68985 70523 74752 75614 75641 76643 76818 76961 78384 79424 80272 81574 83533 85051 85137 85913 86043 86264 86431 89121
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿PKG, CANNULA, 11MM, AUTOMATIC VALVE STOP COCK, W/O TROCAR, P/N 0250080174. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0665-2015 el 24 de diciembre de 2014. Peligro reportado: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter