Computed Tomography X-Ray System Brilliance Workspace Portal, Model #728269 Philips Medical Systems, Highland Heights, OH. The Brilliance Workspace Portal is a component of the Computed Tomography X-Ray System that is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. The portal (together with the scanner) provides image processing and display through software applications that process, analyze, display, quantify and interpret medical images/data.

Publicado el 23 de enero de 2013 · Inicio del retiro: 11 de septiembre de 2012

Resumen

FDA registra el retiro Z-0675-2013 de dispositivos médicos de Clase II, publicado el 23 de enero de 2013 e iniciado el 11 de septiembre de 2012. Empresa responsable: Philips Medical Systems (Cleveland) Inc; Estado: Finalizado; Cantidad reportada: 108 units; Lote: Model #728269; Serial #s: 79906, 79921, 79927, 79937, 80007, 80026, 80031, 80040, 80058, 80062, 80066, 80069, 80087, 80110, 80121, 80155, 80167, 80168, 80169, 80174, 80192, 80193, 80195, 80201, 80208, 80209, 80219, 80227, 80228, 80230, 80252, 80258, 80266, 80280, 80289, 80293, 80294, 80315, 80322, 80328, 80333, 80340, 80341, 80348, 80360, 80375, 80385, 80387, 80394, 80398, 80416, 80421, 80422, 80429, 80447, 80451, 80460, 80474, 80482, 80484, 80485, 80492, 80497, 80505, 80516, 80578, 80585, 80591, 80629, 80646, 80658, 80681, 80704, 80709, 80711, 80720, 80723, 80743, 80749, 80751, 80831, 80835, 80866, 80878, 80908, 80909, 80911, 80914, 80916, 80930, 80931, 80944, 80974, 80992, 80999, 81023, 81025, 81034, 81045, 81071, 81099, 81101, 81107, 81117, 81126, 81136, 95382 & 95592. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Philips was notified that when performing post processing with the Brilliance Workspace Portal platform with software versions 2.6.1.4, the coronary arteries in the rendered image are segmented automatically. However, the center line is not extracted or highlighted upon hovering and making it impossible to activate the vessel and choose the appropriate name (label) from the drop-down menu.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Philips Medical Systems (Cleveland) Inc
Distribuidor
Worldwide Distribution-USA including the states of AZ, CA, CO, IL, IL, IN, MA, ME, MI, MO, NC, NJ, NY, OR, TN, TX, VA and WI and the countries of Austria, Canada, China, France, India, Ireland, Italy, Japan, Netherlands, Norway, Spain, Switzerland, Thailand and United Kingdom.
Lote
Model #728269; Serial #s: 79906, 79921, 79927, 79937, 80007, 80026, 80031, 80040, 80058, 80062, 80066, 80069, 80087, 80110, 80121, 80155, 80167, 80168, 80169, 80174, 80192, 80193, 80195, 80201, 80208, 80209, 80219, 80227, 80228, 80230, 80252, 80258, 80266, 80280, 80289, 80293, 80294, 80315, 80322, 80328, 80333, 80340, 80341, 80348, 80360, 80375, 80385, 80387, 80394, 80398, 80416, 80421, 80422, 80429, 80447, 80451, 80460, 80474, 80482, 80484, 80485, 80492, 80497, 80505, 80516, 80578, 80585, 80591, 80629, 80646, 80658, 80681, 80704, 80709, 80711, 80720, 80723, 80743, 80749, 80751, 80831, 80835, 80866, 80878, 80908, 80909, 80911, 80914, 80916, 80930, 80931, 80944, 80974, 80992, 80999, 81023, 81025, 81034, 81045, 81071, 81099, 81101, 81107, 81117, 81126, 81136, 95382 & 95592.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Computed Tomography X-Ray System Brilliance Workspace Portal, Model #728269 Philips Medical Systems, Highland Heights, OH. The Brilliance Workspace Portal is a component of the Computed Tomography X-Ray System that is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. The portal (together with the scanner) provides image processing and display through software applications that process, analyze, display, quantify and interpret medical images/data. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0675-2013 el 23 de enero de 2013. Peligro reportado: Philips was notified that when performing post processing with the Brilliance Workspace Portal platform with software versions 2.6.1.4, the coronary arteries in the rendered image are segmented automatically. However, the center line is not extracted or highlighted upon hovering and making it impossible to activate the vessel and choose the appropriate name (label) from the drop-down menu.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter