Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.
Publicado el 21 de septiembre de 2022 · Inicio del retiro: 30 de agosto de 2022
Resumen
FDA registra el retiro Z-1744-2022 de dispositivos médicos de Clase II, publicado el 21 de septiembre de 2022 e iniciado el 30 de agosto de 2022. Empresa responsable: Philips Medical Systems (Cleveland) Inc; Estado: En curso; Cantidad reportada: 169 systems; Lote: Model 870231: Software version 16.0.4, UDI (01)00884838099081(10)16.0.4.60330; Software version 16.4.2, UDI (01)0084838103054(10)16.4.2.60330, UDI (01)00884838102965(10)16.4.2.60330, UDI (01)00884838102972(10)16.4.2.60315, UDI (01)00884838102989(10)16.4.2.60330; UDI (01)00884838103191(10)16.4.2.60330; and Software version 16.4.3, UDI (01)00884838102965(10)16.4.3.60360. Model 870237: Software version 16.4.6, UDI (01)00884838103566(10)16.4.6.21197; and Software version 18.0.1, UDI (01)00884838103566(10)18.0.1.2149 and (01)00884838103566(10)18.0.1.21197. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Philips Medical Systems (Cleveland) Inc
- Distribuidor
- Distribution was made to CA, FL, GA, IA, IL, MA, MD, ME, MI, MN, NJ, NM, NV, NY, OH, OK, RI, SC, TN, TX, and WA. There was government distribution and no military distribution. Foreign distribution was made to Australia, Austria, Belgium, China, France, Germany, Italy, Japan, Jordan, Lebanon, Poland, Portugal, Romania, Russian Federation, Singapore, Spain, Switzerland, Taiwan, Thailand, and United Kingdom.
- Lote
- Model 870231: Software version 16.0.4, UDI (01)00884838099081(10)16.0.4.60330; Software version 16.4.2, UDI (01)0084838103054(10)16.4.2.60330, UDI (01)00884838102965(10)16.4.2.60330, UDI (01)00884838102972(10)16.4.2.60315, UDI (01)00884838102989(10)16.4.2.60330; UDI (01)00884838103191(10)16.4.2.60330; and Software version 16.4.3, UDI (01)00884838102965(10)16.4.3.60360. Model 870237: Software version 16.4.6, UDI (01)00884838103566(10)16.4.6.21197; and Software version 18.0.1, UDI (01)00884838103566(10)18.0.1.2149 and (01)00884838103566(10)18.0.1.21197.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAClonidine Hydrochloride Tablets, USP, 0.3 mg, 100-count bottle, Rx Only, Manufactured by: UNICHEM LABORATORIES28 de septiembre de 2022
- FDAZAP-X Radiosurgery System21 de septiembre de 2022
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray system14 de agosto de 2019
- FDABrilliance 64 Model # 728231, computed tomography x-ray system14 de agosto de 2019
- FDABrightView XCT , Model Number 88248231 de julio de 2019
- FDAIQon Spectral CT (Model No. 728332), Software Version 4.7.0 or 4.7.2 Product Usage: Computed Tomography X-ray7 de noviembre de 2018
Preguntas sobre este retiro
¿Pinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1744-2022 el 21 de septiembre de 2022. Peligro reportado: When computing a radiation dose in the system, the exported dose information is incorrect when there is more than one beam attached to the prescription and certain options are selected on the Edit Prescription screen.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter