Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0676-2015

PKG, DEBANEY FORCEPS, P/N 0250080314. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Publicado el 24 de diciembre de 2014 · Inicio del retiro: 17 de noviembre de 2014

Resumen

FDA registra el retiro Z-0676-2015 de dispositivos médicos de Clase II, publicado el 24 de diciembre de 2014 e iniciado el 17 de noviembre de 2014. Empresa responsable: Stryker Endoscopy; Estado: Finalizado; Cantidad reportada: 180,573 units; Lote: 143177 143178 246074 344739 444155 644081 645995 647106 647107 647111 647115 647116 647119 647121 647122 647123 647124 647125 647126 846997 940698 940700 940701 940702 940706 940707 940708 940709 940713 940714 940715 940717 940721 940722 940723 940724 943750 943862 1045977 1045978 1045979 1045980 1046707 1141928 1141929 1143434 1143435 1144968 1145466 1145467 1240129 1240549 1242018 1242851 1242852 1243860 1243861 1244476 1244788 1245555 1245556 1245969 1246404 1246406 1340287 1340723 1340724 1342682 1342842 1342843 1342844 1343524 1343525 1343989 1343990 1344458 1346205 1440786 1441077 1441462 1442554 1443205 644082. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Stryker Endoscopy
Distribuidor
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Lote
143177 143178 246074 344739 444155 644081 645995 647106 647107 647111 647115 647116 647119 647121 647122 647123 647124 647125 647126 846997 940698 940700 940701 940702 940706 940707 940708 940709 940713 940714 940715 940717 940721 940722 940723 940724 943750 943862 1045977 1045978 1045979 1045980 1046707 1141928 1141929 1143434 1143435 1144968 1145466 1145467 1240129 1240549 1242018 1242851 1242852 1243860 1243861 1244476 1244788 1245555 1245556 1245969 1246404 1246406 1340287 1340723 1340724 1342682 1342842 1342843 1342844 1343524 1343525 1343989 1343990 1344458 1346205 1440786 1441077 1441462 1442554 1443205 644082
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿PKG, DEBANEY FORCEPS, P/N 0250080314. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0676-2015 el 24 de diciembre de 2014. Peligro reportado: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter