Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10, Model Number: 722010. 2. Allura Xper FD10/10, Model Number: 722011. 3. Allura Xper FD20, Model Number: 722012. 4. Allura Xper FD20 Biplane, Model Number: 722013. 5. Allura Xper FD10 OR Table, Model Number: 722022. 6. Allura Xper FD20 OR Table, Model Number: 722023. 7. Allura Xper FD20 Biplane OR Table, Model Number: 722025.
Publicado el 18 de diciembre de 2024 · Inicio del retiro: 11 de noviembre de 2024
Resumen
FDA registra el retiro Z-0676-2025 de dispositivos médicos de Clase II, publicado el 18 de diciembre de 2024 e iniciado el 11 de noviembre de 2024. Empresa responsable: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.; Estado: En curso; Cantidad reportada: 2,782 units; Lote: 1. Model Number: 722010. UDI (Serial Numbers): (01)00884838054189(21)164 (164); (01)00884838059030(21)272 (272); (01)00884838059030(21)819 (819); (01)00884838059030(21)828 (828); No UDI (33, 39, 42, 43, 65, 67, 69, 79, 107, 119, 127, 158, 160, 169, 177, 190, 191, 209, 213, 214, 215, 218, 227, 229, 239, 240, 241, 252, 274, 280, 285, 286, 290, 296, 298, 306, 308, 316, 342, 343, 344, 360, 364, 365, 395, 411, 414, 422, 423, 453, 468, 477, 478, 480, 485, 488, 499, 500, 505, 521, 527, 531, 567, 572, 585, 592, 604, 605, 611, 624, 636, 637, 638, 641, 646, 648, 651, 656, 662, 668, 670, 671, 673, 677, 679, 683, 690, 694, 698, 714, 728, 730, 755, 763, 764, 780, 785, 787, 788, 791, 792, 794, 809, 818, 830, 833, 838, 839, 873, 878, 882, 890, 896, 902, 905, 914, 917, 920, 921, 931, 932, 937, 939, 946, 968, 970, 979, 986, 987, 992, 999, 1005, 1009, 1011, 1025, 1028, 1040, 1041, 1042, 1043, 1048, 1064, 1069, 1078, 1082, 1084, 1109, 1112, 1118, 1123, 1126, 1128, 1132, 1144, 1170, 1173, 1176, 1181, 1183, 1209, 1210, 1214, 1230, 1233, 1234, 1235, 1236, 1261, 1263, 1275, 1279, 1283, 1292, 1297, 1442, 601, 654, 1122, 1198, 332, 777, 666, 301, 608, 1081, 426, 924, 577, 265, 1047, 1187, 219, 959, 901, 552, 888, 1140, 64, 935, 1008, 1213, 1015, 1016, 1017, 1018, 374, 524, 1153, 1192, 1194, 841, 964, 996, 233, 1211, 721, 1190, 1188, 806, 907, 50, 418, 71, 400, 1168, 1116, 1093, 53, 370, 596, 45, 712, 1148, 323, 825, 402, 713, 195, 176, 508, 123, 1197, 91, 602, 245, 535, 230, 231, 232, 991, 248, 92, 187, 502, 1124, 537, 1199, 941, 553, 302, 289, 173, 134, 1137, 948, 509, 619, 1034, 447, 707, 708, 1285, 928, 1113, 802, 803, 192, 613, 614, 393, 506, 369, 1206, 264, 699, 759, 722, 1100, 350, 748, 1220, 657, 658, 659, 854, 565, 734, 735, 736, 737, 228, 642, 1054, 490, 391, 425, 513, 898, 424, 105, 993, 732, 580, 1256, 704, 695, 696, 238, 706, 1253, 1327, 1032, 319, 1252, 469, 541, 731, 1019, 1023, 1215, 247, 1191, 856, 51, 971, 603, 1003, 270, 267, 268, 1242, 47, 1089, 831, 530, 1091, 193, 578, 322, 1161, 126, 197, 884, 146, 62, 1095, 273, 29, 63, 412, 346, 622, 371, 740, 715, 635, 1134, 1135, 520, 936, 990, 1002, 911, 49, 978, 586, 1004, 1036, 653, 1086, 1180, 776, 288, 1245, 72, 625, 1221, 1087, 984, 253, 398, 1125, 1269, 773, 80, 380, 224, 222, 1010, 957, 282, 847, 1151, 1083, 1268, 711, 685, 686, 1104, 663, 664, 863, 799, 1224, 1225, 771, 631, 494, 1250, 652, 1106, 383, 52, 1278, 861, 906, 872, 382, 409, 969, 139, 1202, 28, 410, 1050, 915, 544, 299, 769, 540, 1007, 1305, 1306, 612, 1203, 1216, 620, 300, 1058, 600, 384, 90, 811, 1138, 967, 795, 1154, 655, 610, 895, 136, 944, 255, 415, 866, 954, 720, 83, 1141, 86, 88, 246, 801, 498, 1039, 574, 495, 1265, 1177, 1178, 128, 1105, 1266, 97, 703, 526, 1237, 650, 98, 427, 496, 1286, 897, 649, 401, 963, 876, 275, 800, 1012, 102, 1238, 475, 1090, 261, 739, 356, 877, 34, 35, 249, 1196, 1276, 1294, 927, 375, 162, 333, 943, 293, 903, 1111, 797, 416, 955, 1145, 514, 100, 1273, 1157, 860, 942, 961, 287, 1021, 1033, 998, 733, 154, 353, 1150, 832, 474, 471, 962, 1037, 1053, 579, 211, 404, 428, 503, 493, 1120, 837, 406, 351, 528, 559, 217, 529, 58, 566, 588, 589, 590, 591, 593, 594, 891, 949, 1146, 1267, 518, 568, 842, 744, 106, 103, 104, 385, 956, 70, 850, 250, 1097, 904, 75, 199, 328, 182, 294, 77, 751, 810, 1280, 1300, 1077, 858, 362, 1207, 874, 131, 1218, 1130, 972, 710, 934, 587, 645, 855, 178, 790, 466, 94, 276, 952, 548, 115, 997, 116, 753, 606, 320, 114, 181, 1182, 1065, 562, 180, 789, 1067, 1129, 975, 626, 89, 523, 243, 223, 1617, 186, 555, 598, 1119, 702, 321, 338, 137, 1051, 660, 1088, 741, 742, 1029, 1136, 95, 844, 973, 1162, 1000, 394, 142, 561, 277, 269, 304, 558, 132, 804, 1044, 822, 852, 476, 862, 96, 396, 254, 373, 875, 312, 313, 900, 220, 563, 330, 367, 473, 515, 909, 112, 121, 762, 564, 1074, 1075, 20, 124, 357, 807, 834, 1066, 184, 697, 857, 634, 980, 782, 783, 125, 1096, 225, 812, 813, 1055, 913, 1185, 368, 349, 570, 781, 354, 386, 387, 525, 66, 388, 235, 236, 549, 251, 278, 868, 1006, 1060, 516, 51. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems over 10 years of life with LTE kits are experiencing propeller motor clamping bolt and x-ray tube locking bolt breaks. If this happens, it may cause erratic C-arm movements and potential collision with patients or bystanders, or detection of a false collisions and subsequent blocked movement of the c-arm. Movement may also cause image noise and black area on the x-ray image, and possible need to expose the patient to additional unnecessary radiation during re-examination.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Distribuidor
- Nationwide distribution. International distribution to Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Oman, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen.
- Lote
- 1. Model Number: 722010. UDI (Serial Numbers): (01)00884838054189(21)164 (164); (01)00884838059030(21)272 (272); (01)00884838059030(21)819 (819); (01)00884838059030(21)828 (828); No UDI (33, 39, 42, 43, 65, 67, 69, 79, 107, 119, 127, 158, 160, 169, 177, 190, 191, 209, 213, 214, 215, 218, 227, 229, 239, 240, 241, 252, 274, 280, 285, 286, 290, 296, 298, 306, 308, 316, 342, 343, 344, 360, 364, 365, 395, 411, 414, 422, 423, 453, 468, 477, 478, 480, 485, 488, 499, 500, 505, 521, 527, 531, 567, 572, 585, 592, 604, 605, 611, 624, 636, 637, 638, 641, 646, 648, 651, 656, 662, 668, 670, 671, 673, 677, 679, 683, 690, 694, 698, 714, 728, 730, 755, 763, 764, 780, 785, 787, 788, 791, 792, 794, 809, 818, 830, 833, 838, 839, 873, 878, 882, 890, 896, 902, 905, 914, 917, 920, 921, 931, 932, 937, 939, 946, 968, 970, 979, 986, 987, 992, 999, 1005, 1009, 1011, 1025, 1028, 1040, 1041, 1042, 1043, 1048, 1064, 1069, 1078, 1082, 1084, 1109, 1112, 1118, 1123, 1126, 1128, 1132, 1144, 1170, 1173, 1176, 1181, 1183, 1209, 1210, 1214, 1230, 1233, 1234, 1235, 1236, 1261, 1263, 1275, 1279, 1283, 1292, 1297, 1442, 601, 654, 1122, 1198, 332, 777, 666, 301, 608, 1081, 426, 924, 577, 265, 1047, 1187, 219, 959, 901, 552, 888, 1140, 64, 935, 1008, 1213, 1015, 1016, 1017, 1018, 374, 524, 1153, 1192, 1194, 841, 964, 996, 233, 1211, 721, 1190, 1188, 806, 907, 50, 418, 71, 400, 1168, 1116, 1093, 53, 370, 596, 45, 712, 1148, 323, 825, 402, 713, 195, 176, 508, 123, 1197, 91, 602, 245, 535, 230, 231, 232, 991, 248, 92, 187, 502, 1124, 537, 1199, 941, 553, 302, 289, 173, 134, 1137, 948, 509, 619, 1034, 447, 707, 708, 1285, 928, 1113, 802, 803, 192, 613, 614, 393, 506, 369, 1206, 264, 699, 759, 722, 1100, 350, 748, 1220, 657, 658, 659, 854, 565, 734, 735, 736, 737, 228, 642, 1054, 490, 391, 425, 513, 898, 424, 105, 993, 732, 580, 1256, 704, 695, 696, 238, 706, 1253, 1327, 1032, 319, 1252, 469, 541, 731, 1019, 1023, 1215, 247, 1191, 856, 51, 971, 603, 1003, 270, 267, 268, 1242, 47, 1089, 831, 530, 1091, 193, 578, 322, 1161, 126, 197, 884, 146, 62, 1095, 273, 29, 63, 412, 346, 622, 371, 740, 715, 635, 1134, 1135, 520, 936, 990, 1002, 911, 49, 978, 586, 1004, 1036, 653, 1086, 1180, 776, 288, 1245, 72, 625, 1221, 1087, 984, 253, 398, 1125, 1269, 773, 80, 380, 224, 222, 1010, 957, 282, 847, 1151, 1083, 1268, 711, 685, 686, 1104, 663, 664, 863, 799, 1224, 1225, 771, 631, 494, 1250, 652, 1106, 383, 52, 1278, 861, 906, 872, 382, 409, 969, 139, 1202, 28, 410, 1050, 915, 544, 299, 769, 540, 1007, 1305, 1306, 612, 1203, 1216, 620, 300, 1058, 600, 384, 90, 811, 1138, 967, 795, 1154, 655, 610, 895, 136, 944, 255, 415, 866, 954, 720, 83, 1141, 86, 88, 246, 801, 498, 1039, 574, 495, 1265, 1177, 1178, 128, 1105, 1266, 97, 703, 526, 1237, 650, 98, 427, 496, 1286, 897, 649, 401, 963, 876, 275, 800, 1012, 102, 1238, 475, 1090, 261, 739, 356, 877, 34, 35, 249, 1196, 1276, 1294, 927, 375, 162, 333, 943, 293, 903, 1111, 797, 416, 955, 1145, 514, 100, 1273, 1157, 860, 942, 961, 287, 1021, 1033, 998, 733, 154, 353, 1150, 832, 474, 471, 962, 1037, 1053, 579, 211, 404, 428, 503, 493, 1120, 837, 406, 351, 528, 559, 217, 529, 58, 566, 588, 589, 590, 591, 593, 594, 891, 949, 1146, 1267, 518, 568, 842, 744, 106, 103, 104, 385, 956, 70, 850, 250, 1097, 904, 75, 199, 328, 182, 294, 77, 751, 810, 1280, 1300, 1077, 858, 362, 1207, 874, 131, 1218, 1130, 972, 710, 934, 587, 645, 855, 178, 790, 466, 94, 276, 952, 548, 115, 997, 116, 753, 606, 320, 114, 181, 1182, 1065, 562, 180, 789, 1067, 1129, 975, 626, 89, 523, 243, 223, 1617, 186, 555, 598, 1119, 702, 321, 338, 137, 1051, 660, 1088, 741, 742, 1029, 1136, 95, 844, 973, 1162, 1000, 394, 142, 561, 277, 269, 304, 558, 132, 804, 1044, 822, 852, 476, 862, 96, 396, 254, 373, 875, 312, 313, 900, 220, 563, 330, 367, 473, 515, 909, 112, 121, 762, 564, 1074, 1075, 20, 124, 357, 807, 834, 1066, 184, 697, 857, 634, 980, 782, 783, 125, 1096, 225, 812, 813, 1055, 913, 1185, 368, 349, 570, 781, 354, 386, 387, 525, 66, 388, 235, 236, 549, 251, 278, 868, 1006, 1060, 516, 51
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Netherlands
Retiros relacionados
- FDAPhilips Azurion 5 M12. Model Number: 722227.4 de noviembre de 2026
- FDAPhilips Azurion 3 M12. Model Numbers: 722063, 722221.4 de noviembre de 2026
- FDAPhilips Azurion 7 M20. Model Numbers: 722079, 722224, 722282.4 de noviembre de 2026
- FDAPhilips Azurion 5 M20. Model Numbers: 722228, 722281.4 de noviembre de 2026
- FDACardinal Health SMS Elastic Cuff Thumbhook Level 3 Isolation Gown, Full-Back, Product Codes AT6588-2XL,18 de diciembre de 2024
- FDADiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stool18 de diciembre de 2024
Preguntas sobre este retiro
¿Philips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10, Model Number: 722010. 2. Allura Xper FD10/10, Model Number: 722011. 3. Allura Xper FD20, Model Number: 722012. 4. Allura Xper FD20 Biplane, Model Number: 722013. 5. Allura Xper FD10 OR Table, Model Number: 722022. 6. Allura Xper FD20 OR Table, Model Number: 722023. 7. Allura Xper FD20 Biplane OR Table, Model Number: 722025. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0676-2025 el 18 de diciembre de 2024. Peligro reportado: Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems over 10 years of life with LTE kits are experiencing propeller motor clamping bolt and x-ray tube locking bolt breaks. If this happens, it may cause erratic C-arm movements and potential collision with patients or bystanders, or detection of a false collisions and subsequent blocked movement of the c-arm. Movement may also cause image noise and black area on the x-ray image, and possible need to expose the patient to additional unnecessary radiation during re-examination.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter