PKG, GRASPER FORCEPS, DELICATE, FENESTRATED, P/N 0250080322. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.
Publicado el 24 de diciembre de 2014 · Inicio del retiro: 17 de noviembre de 2014
Resumen
FDA registra el retiro Z-0685-2015 de dispositivos médicos de Clase II, publicado el 24 de diciembre de 2014 e iniciado el 17 de noviembre de 2014. Empresa responsable: Stryker Endoscopy; Estado: Finalizado; Cantidad reportada: 180,573 units total; Lote: 43998 44285 740839 840085 844271 845502 942652 943741 944218 944701 945747 945987 1040596 1042092 1042500 1043510 1045652 1141382 1143801 1144307 1144308 1145636 1146675 1240767 1241940 1242644 1244526 1244963 1246365 1340542 1341581 1341981 1342752 1343399 1343699 1346235 1440401. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Stryker Endoscopy
- Distribuidor
- Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
- Lote
- 43998 44285 740839 840085 844271 845502 942652 943741 944218 944701 945747 945987 1040596 1042092 1042500 1043510 1045652 1141382 1143801 1144307 1144308 1145636 1146675 1240767 1241940 1242644 1244526 1244963 1246365 1340542 1341581 1341981 1342752 1343399 1343699 1346235 1440401
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAPKG, BIPOLAR FORCEPS, FENESTRATED, 33CM, P/N 0250080123 Laparoscopic Manual Instruments are intended for24 de diciembre de 2014
- FDAPKG, ENDO METZENBAUM SCISSORS, CURVED, P/N 0250080267. Laparoscopic Manual Instruments are intended for24 de diciembre de 2014
- FDAPKG, ALLIGATOR FORCEPS, P/N 0250080316. Laparoscopic Manual Instruments are intended for cutting, holding,24 de diciembre de 2014
- FDAPKG, 5MM X 33CM, INSERT, FENESTRATED GRASPER, P/N 0250080755 Laparoscopic Manual Instruments are intended for24 de diciembre de 2014
- FDAPKG, BABCOCK FORCEPS, P/N 0250080321. Laparoscopic Manual Instruments are intended for cutting, holding,24 de diciembre de 2014
- FDAPKG, GRASPER FORCEPS, ATRAUMATIC, P/N 0250080323. Laparoscopic Manual Instruments are intended for cutting,24 de diciembre de 2014
Preguntas sobre este retiro
¿PKG, GRASPER FORCEPS, DELICATE, FENESTRATED, P/N 0250080322. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0685-2015 el 24 de diciembre de 2014. Peligro reportado: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter