Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0714-2017

Peripherally Inserted Central Catheters (PICC), UPN H965750141, Catalog # 75-014 The Maximal Barrier Nursing Kits are Tandem Pack Convenience Kits packaged with a NMI PICC and accessories. The BioFlo PICC with ENDEXO Technology and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring

Publicado el 14 de diciembre de 2016 · Inicio del retiro: 8 de junio de 2016

Resumen

FDA registra el retiro Z-0714-2017 de dispositivos médicos de Clase II, publicado el 14 de diciembre de 2016 e iniciado el 8 de junio de 2016. Empresa responsable: Angiodynamics Inc. (Navilyst Medical Inc.); Estado: Finalizado; Cantidad reportada: 48 cartons (boxes of 5); Lote: Lot #: 4941411. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

There is a potential for stylet stiffening that is not accurately represented on the labeling. The label states that product includes a stylet with an outside diameter (OD) measuring 0.014, however, the product may contain a stylet with an outside diameter (OD) measuring 0.016.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Angiodynamics Inc. (Navilyst Medical Inc.)
Distribuidor
Nationwide Distribution to CA, FL, GA, HI, LA, MI, MN, MO, NV, OH & TX
Lote
Lot #: 4941411
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Peripherally Inserted Central Catheters (PICC), UPN H965750141, Catalog # 75-014 The Maximal Barrier Nursing Kits are Tandem Pack Convenience Kits packaged with a NMI PICC and accessories. The BioFlo PICC with ENDEXO Technology and BioFlo PICC with ENDEXO and PASV Valve Technology are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications, nutrients; the sampling of blood; and for power injection of contrast media. Non-Valve lumens are indicated for central venous pressure monitoring fue retirado del mercado?

Sí. FDA publicó el retiro Z-0714-2017 el 14 de diciembre de 2016. Peligro reportado: There is a potential for stylet stiffening that is not accurately represented on the labeling. The label states that product includes a stylet with an outside diameter (OD) measuring 0.014, however, the product may contain a stylet with an outside diameter (OD) measuring 0.016.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter