Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0758-2015

PKG, 3MM INSERT, METZENBAUM, CURVED, LONG SCISSORS, 20CM, P/N 0250282000 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Publicado el 24 de diciembre de 2014 · Inicio del retiro: 17 de noviembre de 2014

Resumen

FDA registra el retiro Z-0758-2015 de dispositivos médicos de Clase II, publicado el 24 de diciembre de 2014 e iniciado el 17 de noviembre de 2014. Empresa responsable: Stryker Endoscopy; Estado: Finalizado; Cantidad reportada: 180,573 units total; Lote: 342970 645484 743722 747490 747891 840091 945208 945712 946001 1040121 1041472 1042482 1042952 1046585 1140928 1143074 1143076 1143121 1145963 1146536 1241511 1244708 1246167 1341512 1343689 1344271 1344682 1440414. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Stryker Endoscopy
Distribuidor
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Lote
342970 645484 743722 747490 747891 840091 945208 945712 946001 1040121 1041472 1042482 1042952 1046585 1140928 1143074 1143076 1143121 1145963 1146536 1241511 1244708 1246167 1341512 1343689 1344271 1344682 1440414
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿PKG, 3MM INSERT, METZENBAUM, CURVED, LONG SCISSORS, 20CM, P/N 0250282000 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0758-2015 el 24 de diciembre de 2014. Peligro reportado: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter