Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.
Publicado el 6 de febrero de 2019 · Inicio del retiro: 21 de diciembre de 2018
Resumen
FDA registra el retiro Z-0763-2019 de dispositivos médicos de Clase I, publicado el 6 de febrero de 2019 e iniciado el 21 de diciembre de 2018. Empresa responsable: Edwards Lifesciences, LLC; Estado: Finalizado; Cantidad reportada: 875 units; Lote: Lot Numbers: 61321241, 61176369, 61176374, 61311580, 61176373, 61321254. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Edwards Lifesciences, LLC
- Distribuidor
- Worldwide Distribution
- Lote
- Lot Numbers: 61321241, 61176369, 61176374, 61311580, 61176373, 61321254
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAGeneric Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery System11 de noviembre de 2026
- FDAEdwards EVOQUE tricuspid delivery system REF:9850TDS Sterile EO20 de mayo de 2026
- FDABrand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016AS Software16 de julio de 2025
- FDABrand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI1616 de julio de 2025
- FDAPlums sold at ALDI, packaged in a 2-pound bag with the brand Rio Duero, EAN# 7804650090304.6 de febrero de 2019
- FDAEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF 151F7 Product Usage: The Swan-Ganz Thermodilution6 de febrero de 2019
Preguntas sobre este retiro
¿Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0763-2019 el 6 de febrero de 2019. Peligro reportado: These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter