Carestream DRX Evolution System (Standard Q) The DRX-Evolution System (Standard Q) is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions. These products are permanently installed diagnostic x-ray systems composed of 4 main components: an operator console, over head tube crane, x-ray tube assembly including a collimator and stationary generator. In addition the device can use an x-ray table and /or wallstandbucky to complete x-ray exposures.
Publicado el 5 de febrero de 2014 · Inicio del retiro: 20 de agosto de 2013
Resumen
FDA registra el retiro Z-0774-2014 de dispositivos médicos de Clase II, publicado el 5 de febrero de 2014 e iniciado el 20 de agosto de 2013. Empresa responsable: Carestream Health Inc.; Estado: Finalizado; Cantidad reportada: 4 units; Lote: Serial Numbers 5049, 5073, 5133, 5162. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Carestream received a report of an exposure being initiated without a patient image captured. The patient was imaged 2 times before an image was captured. The field engineer who visited the site reported the system allowed an exposure without the assignment of a bucky or image receptor.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Carestream Health Inc.
- Distribuidor
- Nationwide Distribution including TX, IL, and GA.
- Lote
- Serial Numbers 5049, 5073, 5133, 5162
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAUltra-speed Dental Film, Catalog Number/REF 834 8658, KODAK and Carestream DENTAL labels --- Carestream Health13 de agosto de 2014
- FDAKODAK DirectView DR 9000 System, Catalog Numbers 1966688 (US and Canada) and 8161937 (outside the US and13 de agosto de 2014
- FDAINSIGHT Dental Film, Catalog Number/REF 811 0785, KODAK and Carestream DENTAL labels --- Carestream Health13 de agosto de 2014
- FDAModel Kodak 2200 Intraoral X-Ray System, Catalog Numbers 5303177, 5303169, 5303151, XR-5154323 -- Carestream12 de marzo de 2014
- FDANatureBox Salsa Spiced Nut Mix; Net. WT: 4 oz. (113 g), Stand up Pouch, Vegan Ingredients: Almonds, Corn chips5 de febrero de 2014
- FDABrilliance iCT, Brilliance 64 and Brilliance Big Bore The Brilliance iCT, Brilliance 64 and Brilliance Big5 de febrero de 2014
Preguntas sobre este retiro
¿Carestream DRX Evolution System (Standard Q) The DRX-Evolution System (Standard Q) is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions. These products are permanently installed diagnostic x-ray systems composed of 4 main components: an operator console, over head tube crane, x-ray tube assembly including a collimator and stationary generator. In addition the device can use an x-ray table and /or wallstandbucky to complete x-ray exposures. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0774-2014 el 5 de febrero de 2014. Peligro reportado: Carestream received a report of an exposure being initiated without a patient image captured. The patient was imaged 2 times before an image was captured. The field engineer who visited the site reported the system allowed an exposure without the assignment of a bucky or image receptor.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter