Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0790-2015

PKG, 3MM INSERT, DORSEY GRASPER, 29CM, P/N 0250282097 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Publicado el 24 de diciembre de 2014 · Inicio del retiro: 17 de noviembre de 2014

Resumen

FDA registra el retiro Z-0790-2015 de dispositivos médicos de Clase II, publicado el 24 de diciembre de 2014 e iniciado el 17 de noviembre de 2014. Empresa responsable: Stryker Endoscopy; Estado: Finalizado; Cantidad reportada: 180,573 units total; Lote: 742567 742873 944138 944717 944818 945995 1040387 1040404 1042105 1042913 1042914 1043535 1043821 1044266 1045473 1046692 1141212 1141216 1142924 1143141 1345284. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Stryker Endoscopy
Distribuidor
Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.
Lote
742567 742873 944138 944717 944818 945995 1040387 1040404 1042105 1042913 1042914 1043535 1043821 1044266 1045473 1046692 1141212 1141216 1142924 1143141 1345284
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿PKG, 3MM INSERT, DORSEY GRASPER, 29CM, P/N 0250282097 Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0790-2015 el 24 de diciembre de 2014. Peligro reportado: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter