Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0816-2015

EPIQ 5 Ultrasound System, EPIQ 5 systems with software versions 1.0.x or 1.1.x with QLAB a2DQ and/or aCMQ plug-ins are affected. Model: EPIQ 5G, EPIQ 5C, EPIQ 5W. Catalog Number: 795204 / 795205 and Part Number: 989605408541 Diagnostic Ultrasound Imaging.

Publicado el 24 de diciembre de 2014 · Inicio del retiro: 5 de noviembre de 2014

Resumen

FDA registra el retiro Z-0816-2015 de dispositivos médicos de Clase II, publicado el 24 de diciembre de 2014 e iniciado el 5 de noviembre de 2014. Empresa responsable: Philips Ultrasound, Inc.; Estado: Finalizado; Cantidad reportada: 88 units (42 in the US and 46 outside US); Lote: SERIAL NUMBERS FOR FCO 79500317-EPIQ5 system: US114B0005, US114B0006, US114C0193, US114C0209, US114C0239, US114C0240, US114C0241, US114C0271, US114C0272, US114C0273, US114C0284, US114C0286, US114C0287, US114C0308, US114C0331, US114C0334, US114C0366, US114C0751, US214C0043, US214C0106, US214C0140, US214C0143, US214C0144, US214C0416, US214C0417, US314C0199, US314C0298, US314C0447, US314C0728, US314C0745, US314C0800, US314C0801, US414C0012, US414C0066, US414C0114, US414C0157, US414C0158, US414C0159, US414C0395, US514C0081, USD13B0662, USD13C0033, USD13C0234, USD13C0290, USD13C0291, USD13C0481, USD13C0525, USD13C0526, USD13C0527, USD13C0572, USD13C0573, USD13C0584, USD13C0709, USN13B0039, USN13C0018, USN13C0182, USN13C0183, USN13C0184, USN13C0186, USN13C0206, USN13C0281, USN13C0282, USN13C0283, USN13C0284, USN13C0285, USN13C0286, USN13C0312, USN13C0361, USN13C0502, USN13C0503, USN13C0580, USN13C0654, USO13C0352, USO13C0354, USO13C0355, USO13C0356, USO13C0368, USO13C0369, USO13C0395, USO13C0396, USO13C0397, USO13C0398, USO13C0405, USO13C0453, USO13C0454, USO13C0596, USO13C0678, and USO13C0685. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection Fraction (EF) may be higher, than the ESV and EF calculated by manual tracing without the use of QLAB. An incorrect EF calculation could lead to misdiagnosis and/or delayed or incorrect therapy.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Philips Ultrasound, Inc.
Distribuidor
Worldwide distribution (for FCO 83000190): US, including the states of AK, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NJ, NY, OH, OR, PA, SC, SD, TN, TX, VA, WA, and WI; and the countries of United Arab Emirates, Austria, Australia, Belgium, Canada, Switzerland, Colombia, Germany, Ecuador, Estonia, Spain, Finland, France, Great Britain, Hong Kong, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, Norway, New Zealand, Philippines, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Turkey, Taiwan, and Vietnam. ********************************************************************************** Worldwide Distribution -- US, including the states of KY, TN, ME, CA, PA, NC, CT, NY, SC, IL, IN, and WI; and the countries of Argentina, Spain, India, Italy, Austria, Germany, Belgium, Canada, Finland, Switzerland, Greece, Malaysia, Sweden, Australia, Netherlands, Mexico, Peru, Taiwan, Japan, France, and Norway.
Lote
SERIAL NUMBERS FOR FCO 79500317-EPIQ5 system: US114B0005, US114B0006, US114C0193, US114C0209, US114C0239, US114C0240, US114C0241, US114C0271, US114C0272, US114C0273, US114C0284, US114C0286, US114C0287, US114C0308, US114C0331, US114C0334, US114C0366, US114C0751, US214C0043, US214C0106, US214C0140, US214C0143, US214C0144, US214C0416, US214C0417, US314C0199, US314C0298, US314C0447, US314C0728, US314C0745, US314C0800, US314C0801, US414C0012, US414C0066, US414C0114, US414C0157, US414C0158, US414C0159, US414C0395, US514C0081, USD13B0662, USD13C0033, USD13C0234, USD13C0290, USD13C0291, USD13C0481, USD13C0525, USD13C0526, USD13C0527, USD13C0572, USD13C0573, USD13C0584, USD13C0709, USN13B0039, USN13C0018, USN13C0182, USN13C0183, USN13C0184, USN13C0186, USN13C0206, USN13C0281, USN13C0282, USN13C0283, USN13C0284, USN13C0285, USN13C0286, USN13C0312, USN13C0361, USN13C0502, USN13C0503, USN13C0580, USN13C0654, USO13C0352, USO13C0354, USO13C0355, USO13C0356, USO13C0368, USO13C0369, USO13C0395, USO13C0396, USO13C0397, USO13C0398, USO13C0405, USO13C0453, USO13C0454, USO13C0596, USO13C0678, and USO13C0685.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿EPIQ 5 Ultrasound System, EPIQ 5 systems with software versions 1.0.x or 1.1.x with QLAB a2DQ and/or aCMQ plug-ins are affected. Model: EPIQ 5G, EPIQ 5C, EPIQ 5W. Catalog Number: 795204 / 795205 and Part Number: 989605408541 Diagnostic Ultrasound Imaging. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0816-2015 el 24 de diciembre de 2014. Peligro reportado: When using the QLAB Auto 2D Quantification (a2DQ) and Auto Cardiac Motion Quantification (aCMQ) applications, the reported End-Systolic Volume (ESV) may be smaller, and the Left Ventricular Ejection Fraction (EF) may be higher, than the ESV and EF calculated by manual tracing without the use of QLAB. An incorrect EF calculation could lead to misdiagnosis and/or delayed or incorrect therapy.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter