Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0829-2013

LPS Diaphyseal Sleeves Product Usage: The LPS Diaphyseal Sleeves are used with the LPS System which is an end-stage revision knee product that allows surgeons to reconstruct severe soft tissue and bony defects.

Publicado el 27 de febrero de 2013 · Inicio del retiro: 4 de enero de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

DePuy Orthopaedics is initiating a voluntary recall of the LPS Diaphyseal sleeve because the product has the potential for fracture at the taper junction. The LPS Diaphyseal Sleeve to Diaphyseal Sleeve Base taper connection may not be sufficient to accommodate potential physiologic loads that may be transferred to the junction during normal gait activities by some patients.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
DePuy Orthopaedics, Inc.
Distribuidor
Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, CO, FL, GA, ID, IL, MA, MD, MI, MN, NC, NM, NV, NY, OH, PA, SC, SD, TN, TX, VA, WA, WI, and WV and the countries of Holland, Australia, Austria, South Africa and Canada including Veteran Administration hospitals.
Lote
Part number: 198720020 and Lot numbers: 140576, 156212, 156213, 156214, 162644, 188411, 202954, 156214A, 202954A, B34GD1000, B4SBN1000, B4SBNA000, B4SBNB000, B4SBNC000, B5RAW1000, B5RAWA000, C3RHJ1000, C74JX1000, C74JXA, E18N61000, E3PKV1000, EF8GB1000, and FL2WB1.
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta