8470-009-040 4.0MM Precision Round, 16CM, single use, Sterile The Stryker MIS (Minimally Invasive Surgery) product offering, is used to remove osseous material during surgical cases. Each cutting accessory consists of: " A shank (B) with a notch (A), that fits in the attachment on one end and is connected to the head of the bur on the other end. " A head (C) connected to the shank (B) and removes material using machined flutes. The head is machined directly on the shank/neck of the bur.
Publicado el 28 de diciembre de 2016 · Inicio del retiro: 30 de noviembre de 2016
Resumen
FDA registra el retiro Z-0871-2017 de dispositivos médicos de Clase II, publicado el 28 de diciembre de 2016 e iniciado el 30 de noviembre de 2016. Empresa responsable: Stryker Instruments Div. of Stryker Corporation; Estado: Finalizado; Cantidad reportada: 1,257 in total; Lote: Lot Number:16038017, Product Number: 8470-009-040. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Stryker Instruments is voluntarily recalling specific lots of sterile MIS Toolsteel Burs because the burs may exhibit signs of corrosion.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Stryker Instruments Div. of Stryker Corporation
- Distribuidor
- Worldwide Distribution - US (nationwide) including states of: AK, AL, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NM, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI and countries of: Canada, Japan, Venlo (Netherlands), France, and UK (United Kingdom).
- Lote
- Lot Number:16038017, Product Number: 8470-009-040
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDASurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090. Used for counting,27 de julio de 2022
- FDANeptune E-SEP 165mm Blade Electrode, Catalog Number 0703-165-001 - Product Usage: The Neptune E-SEP is an6 de mayo de 2020
- FDAColor Cuff Disposable Tourniquet Cuff, 18 inch X 3 inch, 1BLA (Single Bladder), 1PRT(Single Port) Quick11 de diciembre de 2019
- FDAC2 Nerve Monitor, 8 Channel, 5140-508-280, UDI 07613327277319 Product Usage: The C2 Nerve Monitor System is9 de octubre de 2019
- FDA8470-009-030 3.0MM Precision Round, 16CM, single use, Sterile The Stryker MIS (Minimally Invasive Surgery)28 de diciembre de 2016
- FDABD SurePathTM Preservative Collection Vials, catalog numbers 490505, 490527, 491095, 491096, and 491097.28 de diciembre de 2016
Preguntas sobre este retiro
¿8470-009-040 4.0MM Precision Round, 16CM, single use, Sterile The Stryker MIS (Minimally Invasive Surgery) product offering, is used to remove osseous material during surgical cases. Each cutting accessory consists of: " A shank (B) with a notch (A), that fits in the attachment on one end and is connected to the head of the bur on the other end. " A head (C) connected to the shank (B) and removes material using machined flutes. The head is machined directly on the shank/neck of the bur. fue retirado del mercado?
Sí. FDA publicó el retiro Z-0871-2017 el 28 de diciembre de 2016. Peligro reportado: Stryker Instruments is voluntarily recalling specific lots of sterile MIS Toolsteel Burs because the burs may exhibit signs of corrosion.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter