Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0889-2020

CS5/5+ Fastpacks, 125 mL, 20® Res Autotransfusion Device- Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.

Publicado el 5 de febrero de 2020 · Inicio del retiro: 21 de octubre de 2019

Resumen

FDA registra el retiro Z-0889-2020 de dispositivos médicos de Clase II, publicado el 5 de febrero de 2020 e iniciado el 21 de octubre de 2019. Empresa responsable: Haemonetics Corporation; Estado: Finalizado; Cantidad reportada: 1340; Lote: Catalog number/FG Item Number: 0265F-00 1. No specific lots with bowl cracks. 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Potential to develop leaks at the inner core of Cell Saver® 5/5+ and Cell Saver® Elite®/Elite® +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Haemonetics Corporation
Distribuidor
Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, and countries of NZ, AU, IN, PK, PH, HK, VN, MY, ID, SG, CN, AT, DE, FR, GB, IT, CH.
Lote
Catalog number/FG Item Number: 0265F-00 1. No specific lots with bowl cracks. 2. Missing instructions to check for cracks in IFU All products/lost manufactured between October 2014 and October 2019.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿CS5/5+ Fastpacks, 125 mL, 20® Res Autotransfusion Device- Product Usage: its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma. fue retirado del mercado?

Sí. FDA publicó el retiro Z-0889-2020 el 5 de febrero de 2020. Peligro reportado: Potential to develop leaks at the inner core of Cell Saver® 5/5+ and Cell Saver® Elite®/Elite® +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter