FDAFDA · Alimentos y medicamentos·N.º oficial Z-0920-2013
Integra CUSA excel. CEM Nosecone for CUSA EXcel System. The CEM Nosecone is intended for use CUSA Excel/CUSA Excel + Ultrasonic Surgical Aspiration System, CUSA handpiece, and Covidien Force FX Electrosurgical Generator. Used in surgical procedures where combined ultrasonic dissection and electrosurgical coagulation is desired, either simultaneously or independently.
Publicado el 20 de marzo de 2013 · Inicio del retiro: 7 de enero de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
There is a potential for erosion of the CUSA EXcel tip used with the Nosecone. Specifically, the potential for tip erosion occurs when using a CUSA CEM Nosecone if the side of the tip, rather than the annulus of the tip, is in contact with the tissue. This erosion can potentially lead to tip breakage. There were no reports of a patient injury associated with these complaints.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Integra LifeSciences Corp.
- Distribuidor
- Worldwide Distribution - USA and the countries of Austria, Australia, Bangladesh, Belgium, Bulgaria, Brazil, Switzerland, Canada, China, Costa Rica, Czech Republic, Germany, Dominican Republic, Egypt, Emirates, Spain, Finland, France, Great Britian, Hong Kong, Hungary, Indonesia, Ireland, Israel, India, Italy, Japan, Korea (South), Kuwait, Lebanon, Sri Lanka, Latvia, Macedonia, Mexico, Malaysia, Netherlands, Philippines, Pakistan, Poland, Portugal, Qatar, Serbia, Saudi Arabia, Singapore, Slovenia, Thailand, Tunisia, Taiwan, United Arab, Viet Nam and South Africa.
- Lote
- CEM Nosecone Catalogue No. C6623 Lot numbers: 1092855, 1093129, 1100436, 1100437, 1101217, 1101328, 1103393, 1104265, 1105067, 1110311, 1110892, 1112182, 1112183, 1114112, 1114655, 1115225, 1115499, 1115500, 1120771, 1121141, 1121142, 1121383, 1121627, 1122258, 1122259, 1123049 and 1130710. CEM Nosecone, Catalogue No. C6636: 1091431, 1091679, 1091680, 1093475, 1095103, 1095416, 1095417, 1100242, 1100434, 1100435, 1101094, 1101218, 1101219, 1103104, 1103395, 1103849, 1103849, 1103850, 1103852, 1103917, 1103918, 1103922, 1104257, 1104263, 1105071, 1110475, 1110476, 1110978, 1111490, 1111493, 1112187, 1112188, 1112212, 1113179, 1113180, 1113663, 1114357, 1114358, 1114659, 1114660, 1115039, 1115502, 1115504, 1115505, 1120776, 1120777, 1121167, 1121168, 1121169, 1121565, 1121566, 1121567, 1122344, 1123051, 1123052, 1123508, 1123511 and 1124053.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States