FDAFDA · Alimentos y medicamentos·N.º oficial Z-0925-2015
(R)Cable remote control JUPITER, Material no.1219655, Serial number: to (incl.) 101998119. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.
Publicado el 14 de enero de 2015 · Inicio del retiro: 3 de diciembre de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Trumpf Medical Systems, Inc.
- Distribuidor
- Worldwide Distribution - USA including AL, AZ, CA, CT, FL, GA, IL, KS, LA, MD, MI, MK, MS, NC, NH, NJ, NY, OH, OK, PA, RI, SC, TX, VA, WA, WI and Internationally to Canada.
- Lote
- The operating tables which are operated with the affected remote controls (all serial numbers): Mobile column JUPITER System U, Material no. 4500113; Mobile column JUPITER System Miyabi U, Material no. 1574977; Operating table column JUPITER SM 360 U, Material no. 1389824; Operating table column JUPITER SF U OCE, Material no. 1622774; Stationary column JUPITER System U, Material no. 4500112; FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277; Floor mounting column JUPITER 360 U, Material no.1389826; Operating table JUPITER UNIVERSAL U, Material no. 4500170; OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States