Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0971-2024

Atlan, Gas Machine for Anesthesia or Analgesia, Catalog numbers 8621500 (Atlan A350) and 8621600 (Atlan A350 XL)

Publicado el 7 de febrero de 2024 · Inicio del retiro: 17 de noviembre de 2023

Resumen

FDA registra el retiro Z-0971-2024 de dispositivos médicos de Clase II, publicado el 7 de febrero de 2024 e iniciado el 17 de noviembre de 2023. Empresa responsable: Draegerwerk Ag & Co. Kgaa; Estado: En curso; Cantidad reportada: 50 imported to US, not distributed to end users; Lote: UDI-DI: 04048675556176 (Atlan A350) 04048675556183 (Atlan A350 XL) Lots: ASSH-0220 ASSH-0221 ASSH-0222 ASSH-0223 ASSH-0224 ASSH-0225 ASSH-0226 ASSH-0227 ASSH-0228 ASSH-0229 ASSH-0230 ASSH-0231 ASSH-0232 ASSH-0233 ASSH-0234 ASSH-0235 ASSH-0236 ASSH-0237 ASSH-0238 ASSH-0239 ASSH-0240 ASSH-0241 ASSH-0242 ASSH-0243 ASSH-0244 ASSH-0245 ASSH-0246 ASSH-0247 ASSH-0248 ASSH-0249 ASSH-0250 ASSH-0251 ASSH-0252 ASSH-0253 ASSH-0254 ASSL-0347 ASSL-0347 ASSL-0349 ASSL-0350 ASSL-0351 ASSL-0352 ASSL-0353 ASSL-0354 ASSL-0355 ASSL-0356 ASSL-0357 ASSL-0358 ASSL-0359 ASSL-0360 ASSL-0361. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Internal backup battery may fail spontaneously while the Atlan is being operated without AC mains supply, resulting in unexpected shutdown of the device while running on batteries.

Qué hacer

Voluntary: Firm initiated · E-Mail

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Draegerwerk Ag & Co. Kgaa
Distribuidor
No devices were distributed to end-users (customers) in the US. All devices are currently under the direct control of Draeger, Inc.
Lote
UDI-DI: 04048675556176 (Atlan A350) 04048675556183 (Atlan A350 XL) Lots: ASSH-0220 ASSH-0221 ASSH-0222 ASSH-0223 ASSH-0224 ASSH-0225 ASSH-0226 ASSH-0227 ASSH-0228 ASSH-0229 ASSH-0230 ASSH-0231 ASSH-0232 ASSH-0233 ASSH-0234 ASSH-0235 ASSH-0236 ASSH-0237 ASSH-0238 ASSH-0239 ASSH-0240 ASSH-0241 ASSH-0242 ASSH-0243 ASSH-0244 ASSH-0245 ASSH-0246 ASSH-0247 ASSH-0248 ASSH-0249 ASSH-0250 ASSH-0251 ASSH-0252 ASSH-0253 ASSH-0254 ASSL-0347 ASSL-0347 ASSL-0349 ASSL-0350 ASSL-0351 ASSL-0352 ASSL-0353 ASSL-0354 ASSL-0355 ASSL-0356 ASSL-0357 ASSL-0358 ASSL-0359 ASSL-0360 ASSL-0361
Clasificación de riesgo
Class II
Área (FDA)
device
País
Germany

Retiros relacionados

Preguntas sobre este retiro

¿Atlan, Gas Machine for Anesthesia or Analgesia, Catalog numbers 8621500 (Atlan A350) and 8621600 (Atlan A350 XL) fue retirado del mercado?

Sí. FDA publicó el retiro Z-0971-2024 el 7 de febrero de 2024. Peligro reportado: Internal backup battery may fail spontaneously while the Atlan is being operated without AC mains supply, resulting in unexpected shutdown of the device while running on batteries.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · E-Mail