Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0978-2014

DMLC IV-ERGO Product Usage: Intended for use with rotating gantry linear accelerators to conform radiation dose delivery to geometrical volumes of specific shape containing pathology to be treated so that adjacent non-diseased tissues are spared to the extent possible.

Publicado el 5 de marzo de 2014 · Inicio del retiro: 1 de mayo de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

It is possible for an intermittent error to occur in the gantry sensor of the DMLC IV Systems

Qué hacer

Voluntary: Firm initiated · Other

Detalles del retiro

Fabricante
Elekta, Inc.
Distribuidor
US Nationwide Distribution in the states of CT, FL, GA, IA, MO, NY, ND, OK, RI, TX, VA, WA
Lote
Identifier : 116001, 116012, 116044, 116055, 116014, 116052, 116023, 116056, 116053, 116043, 116047, 116051, 116122, 116063, 116068, 116013, 116046
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta