FDAFDA · Alimentos y medicamentos·N.º oficial Z-0988-2019
Maquet XS Flat Screen Monitor Holder, PWD50+DF XS32 K3 SPE, Configuration Part Number ARD268400140A Product Usage: This product is designed to support one or two medical grade flat screen(s) alone or in association with surgical lights.
Publicado el 13 de marzo de 2019 · Inicio del retiro: 8 de enero de 2019
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The affected flat screen monitor holders might not be compliant with grounding requirements. As a consequence, there is a risk of mild to severe/fatal electric shock under certain conditions while using the flat screen monitor holder with voltage of 100-220V.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- GETINGE US SALES LLC
- Distribuidor
- The products were distributed to the following US states: AZ and TX.
- Lote
- Serial Numbers 1 2 3 4 5 6
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States