Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-ray system

Publicado el 3 de abril de 2013 · Inicio del retiro: 8 de marzo de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed. Under certain preconditions, executing roadmap OGP (organ program) in the DSA overlay mode may lead to imprecise registration of the DSA mask image with the roadmap image displayed on the live monitor.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Siemens Medical Solutions USA, Inc
Distribuidor
Nationwide Distribution including AL, CA, CO, FL, IA, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, ND, NJ, NY, OH, PA, SC, TX, UT, VA, WA, and WI and to Puerto Rico.
Lote
Model numbers 10094135, 10094137, 10094139, 10094141 and 10280959.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States