FDAFDA · Alimentos y medicamentos·N.º oficial Z-0998-2014

Medtronic Select CAP Arterial Cannulae. Vented, models 78818 (18 Fr.), 78820 (20 Fr.), 78822 (22 Fr.). Non-Vented Cannulae, model 78920 (20 Fr.), 78922 (22 Fr.), 78924 (24 Fr.). Sterilized using Ethylene Oxide. Non-pyrogenic. The Cannula consists of a curved, angled or beveled tip with or without flange that is permanently attached to a molded, clear, flexible PVC tapered body or wirebound PVC body. The cannula body features a tip orientation line to indicate direction of the cannula tip during cannulation. The proximal end of the cannula includes a connection site or a molded 0.95 cm (3/8 in) barbed connector with or without a molded plastic vent plug or peel cap. The vent plug or peel cap allows air to be vented from the cannula before connection to the perfusion line. The proximal end features a pressure monitoring line with a 3-way stopcock. These cannulae are intended for use in perfusion of the ascending aorta during short term (6 hours or less) cardiopulmonary bypass.

Publicado el 26 de febrero de 2014 · Inicio del retiro: 30 de enero de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Medtronic is recalling all models of its Select 3D and Select CAP Arterial Cannula product families due to reports of the cannula body splitting near the suture collar.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Medtronic Inc. Cardiac Rhythm Disease Management
Distribuidor
Worldwide distribution: US (Nationwide) including Puerto Rico, and Internationally to: Algeria, Argentina, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, Cyprus, El Salvador, France, Georgia, Germany, Greece, Hungary, India, Ireland, Italy, Japan, Jordan, Kuwait, Netherlands, Oman, Paraguay, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom, Venezuela, and Yemen.
Lote
Model 78818, Lot Numbers: 2011021826, 2011070534, 2011090245, 2011110207, 2011111844. Model 78820, Lot Numbers: 2011062305, 2011072337. Model 78822, Lot Number: 2011011178 Model 78920, Lot Numbers: 2011050717, 2011082464, 2011120308, 2012070891, 2012071268, 2013081135, 2013121551, 201106C038. Model 78922, Lot Numbers: 2011011187, 2011020121, 2011020785, 2011021532, 2011070801, 2011081266, 2011090483, 2012012160, 2012021436, 2012021695, 2012071735, 2013010416, 2013030273, 2013050807, 2013051036, 2013070174, 2013091410, 2013101028, 201311C767. Model 78924, Lot Numbers: 2011041201, 2011041542, 2011080058, 2011080606, 2011081545, 2011081637, 2011110914, 2012031176, 2012051794, 2012100256, 2012111224, 2013030583, 2013050865, 2013060210.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States