FDAFDA · Alimentos y medicamentos·N.º oficial Z-1024-2016
SENSE Body Coil 1.5T with identification; used in the ACS NT, Intera 1.5T and Achieva 1.5T MR system
Publicado el 9 de marzo de 2016 · Inicio del retiro: 6 de julio de 2015
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Wrong positioning of the coil cables can create unintended resonances via the patient or elevated coil cable temperature due to coupling to the QBC. The combined use of the Sense coil, high SAR level scanning and placement of the cables contrary to instructions for use can lead to patient burns.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Philips Electronics North America Corporation
- Distribuidor
- US: Nationwide OUS: Canada, Afghanistan Albania Algeria Andorra Argentina Australia Austria Azerbaijan Bahamas Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Botswana Brazil Burkina Faso Chile China Colombia Columbia Congo, Democratic Costa Rica Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Ethiopia Finland France Georgia Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Japan Jersey Jordan Kazakhstan Kenya Korea, Republic of Kuwait Kyrgyzstan Latvia Lebanon Libya Lithuania Luxembourg Malaysia Malta Martinique Mauritius Mexico Monaco Mongolia Morocco Mozambique Nepal Netherlands New Caledonia New Zealand Norway Oman Pakistan Palestinian territory, occupied Panama Papua New Guinea Peru Philippines Poland Portugal Puerto Rico Romania Russian Federation Rwanda Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sudan Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Venezuela Viet Nam Yemen Zimbabwe
- Lote
- 4522 131 5575x 4522 132 1985x 4522 132 7619x 4598 000 5187x 4598 000 5189x 4598 003 59051
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States