FDAFDA · Alimentos y medicamentos·N.º oficial Z-1047-2014
ANGIO PACK #2; REF: DYNJ39185; Packaged in Mexico by: Medline Industries, Inc., Mundelein, IL 60060; 1-800-Medline; www.medline.com Surgical kit used to supply tools required to perform angiography medical procedures.
Publicado el 5 de marzo de 2014 · Inicio del retiro: 14 de agosto de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Guidewire .035x150 3MMJ TCFC, item number 88241, has a potential for the coating to flake off of the wire. The guidewire is used in a number of surgical convenience kits assembled and marketed by Medline Industries, Inc.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Medline Industries, Inc.
- Distribuidor
- Nationwide Distribution: US to states of: AL, CA, FL, IL, LA, OH and VA.
- Lote
- Pack Number: DYNJ39185; Lots: 13EB2046, 13FB1553, 13GB2605
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.