Puritan Bennett 980 Ventilator System, Model No. PB980 Ventilator (980xxxxxxxx), The Puritan BennettTM 980 Ventilator System is designed for use on Neonatal (NICU) through Adult patient populations who require respiratory support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb). It is suitable for service in a hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilator support, delivered invasively or noninvasively, to patients who require the following types of ventilator support: "Positive Pressure Ventilation, delivered invasively (via endotracheal tube or trach tube) or non-invasively (via mask or nasal prongs) "Assist/ Control, SIMV or Spontaneous modes of ventilation.
Publicado el 11 de febrero de 2015 · Inicio del retiro: 12 de enero de 2015
Resumen
FDA registra el retiro Z-1058-2015 de dispositivos médicos de Clase II, publicado el 11 de febrero de 2015 e iniciado el 12 de enero de 2015. Empresa responsable: Covidien LP (formerly Nellcor Puritan Bennett Inc.); Estado: Finalizado; Cantidad reportada: 418 units; Lote: All Puritan Bennett 980 ventilators. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Covidien is issuing a voluntary field action for all Puritan Bennett 980 ventilators due to occasional GUI transient resets that last approximately 30 seconds.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Covidien LP (formerly Nellcor Puritan Bennett Inc.)
- Distribuidor
- Worldwide Distribution - USA (nationwide) and Internationally to Canada.
- Lote
- All Puritan Bennett 980 ventilators
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAErythromycin Pledgets USP, 2%, 60-count jar,Rx only, Manufactured by Ei LLC, Kannapolis, NC 28083, Marketed by18 de febrero de 2015
- FDACompounded Plaquex, intravenous injection, 50 ML vial. Each ml contains: Phosphatidyl Choline 50 mg, Sodium18 de febrero de 2015
- FDASoybean Sprouts, net wt. 1.5 lbs., sold under brand Winter Blossom Bean Sprouts brand. Product is packaged in18 de febrero de 2015
- FDAKYPHON EXPRESS, Osteo Introducer System, Diamond and Bevel, Size 2, REF T34A, QTY: 1EA, Rx only, STERILE R.11 de febrero de 2015
- FDABD 10 mL Syringe BD Luer-Lok Tip with Blunt Plastic Cannula, Not for Skin Injection ; Catalog #303348, BD 1011 de febrero de 2015
- FDAMobileDiagnost wDR motorized portable diagnostic X-ray systems.11 de febrero de 2015
Preguntas sobre este retiro
¿Puritan Bennett 980 Ventilator System, Model No. PB980 Ventilator (980xxxxxxxx), The Puritan BennettTM 980 Ventilator System is designed for use on Neonatal (NICU) through Adult patient populations who require respiratory support or mechanical ventilation and weigh a minimum of 0.3 kg (0.66 lb). It is suitable for service in a hospital (institutions) and intra-hospital transport to provide continuous positive pressure ventilator support, delivered invasively or noninvasively, to patients who require the following types of ventilator support: "Positive Pressure Ventilation, delivered invasively (via endotracheal tube or trach tube) or non-invasively (via mask or nasal prongs) "Assist/ Control, SIMV or Spontaneous modes of ventilation. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1058-2015 el 11 de febrero de 2015. Peligro reportado: Covidien is issuing a voluntary field action for all Puritan Bennett 980 ventilators due to occasional GUI transient resets that last approximately 30 seconds.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter