Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1071-2024

Single Use Distal Cover MAJ-2315 Used with: EVIS EXERA III Duodenovideoscope TJF-Q190V- designed to be attached to Olympus duodenovideoscopes to cover the distal end of the insertion tube and around the forceps elevator

Publicado el 21 de febrero de 2024 · Inicio del retiro: 8 de diciembre de 2023

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Single Use Distal Cover MAJ-2315 may unexpectedly detach from the duodenoscope during use result include the risk of mucosal injury, tissue damage, bleeding, perforation, or obstruction due to retained distal cover in the gastrointestinal tract that would require urgent medical intervention for removal, and/or burns from the use an uncovered duodenoscope distal end. Detachment of a distal cover in the oral cavity may result in aspiration, airway obstruction, or respiratory distress and would require immediate medical intervention for removal.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Olympus Corporation of the Americas
Distribuidor
US Nationwide distribution.
Lote
UDI: 04953170441271 All lot numbers
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States