VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), Product Code 6802445. Intended for use in the measurement of a variety of chemistry analytes.

Publicado el 24 de abril de 2013 · Inicio del retiro: 7 de febrero de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Ortho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS¿ Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal evaluation in its manufacturing facility, it was discovered that an extension spring had insufficient force. If the extension spring does not provide sufficient force, MicroSlides may not be properly positioned within the MicroSlide incubator which has the potential to cause imprecise results. Extension springs are just one component in the slide alignment assembly. OCD's evaluation determined that the probability of an out of specification extension spring alone adversely affecting results is unlikely.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Ortho-Clinical Diagnostics
Distribuidor
Worldwide Distribution- USA (nationwide) including Puerto Rico and countries of: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Republic of Panama, Singapore, Spain, and Venezuela.
Lote
Serial Numbers 46000108 through 46000217
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States