The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA alkaline batteries. When powered by batteries, The GemStar Pump is ideal for ambulatory patients.
Publicado el 24 de abril de 2013 · Inicio del retiro: 21 de marzo de 2013
Resumen
FDA registra el retiro Z-1117-2013 de dispositivos médicos de Clase II, publicado el 24 de abril de 2013 e iniciado el 21 de marzo de 2013. Empresa responsable: Hospira Inc.; Estado: Finalizado; Cantidad reportada: 1) List Number 13086: 286 pumps; 3) List Number 13087: 45,376 pumps; 4) List Number 13088: 23,492 pumps; Lote: List Numbers: 13086, 13087, 13088. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
During infusions of 2.0 mL/hr the motor assembly may rotate backwards capturing additional medication resulting in over-infusion.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Hospira Inc.
- Distribuidor
- Worldwide Distribution - USA (nationwide) and Internationally to Australia, New Zealand, China, Hong Kong, Taiwan, Japan, Korea, Philippines, Thailand, Malaysia, Canada, Austria, Bahrain, Belgium, Croatia, Denmark, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Iceland, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Netherlands, Norway, Oman, Portugal, Qatar, Romania, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UK, United Arab Emirates, Brazil, Chile and Colombia.
- Lote
- List Numbers: 13086, 13087, 13088
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAInfant 25% DEXTROSE Injection, USP, 2.5 g (250 mg/mL), 10 mL Unit of Use Single-dose Syringe per carton, Rx31 de mayo de 2017
- FDAMinibore Extension Set, 59 Inch with Spin Lock Collar, Non-DEHP, Latex-Free Intended use: For the18 de enero de 2017
- FDAFentanyl Citrate Injection, USP, 100 mcg Fentanyl/ 2mL, Single Dose, 50 mcg/mL, Rx only, Hospira, Inc., Lake30 de noviembre de 2016
- FDASymbiq Two Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids. List number16 de noviembre de 2016
- FDAThe GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in24 de abril de 2013
- FDAClinical Chemistry Phenobarbital, list number 1E0824 de abril de 2013
Preguntas sobre este retiro
¿The GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in the hospital, or anywhere electronic infusion is required. The GemStar Pump can be powered by AC mains adaptor, rechargeable battery pack, docking station, or two disposable AA alkaline batteries. When powered by batteries, The GemStar Pump is ideal for ambulatory patients. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1117-2013 el 24 de abril de 2013. Peligro reportado: During infusions of 2.0 mL/hr the motor assembly may rotate backwards capturing additional medication resulting in over-infusion.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter