Azurion 3 with a Certeray generator -To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722221 (2) 722222 (3) 722280
Publicado el 28 de febrero de 2024 · Inicio del retiro: 6 de febrero de 2024
Resumen
FDA registra el retiro Z-1176-2024 de dispositivos médicos de Clase II, publicado el 28 de febrero de 2024 e iniciado el 6 de febrero de 2024. Empresa responsable: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.; Estado: En curso; Cantidad reportada: 14 units; Lote: OUS: (1) Model: 722221 SN: 66 (01)00884838099203(21)66; SN: 91 (01)00884838099203(21)91; SN: 104 (01)00884838099258(21)104, (2) Model: 722222 SN: UDI: 165 (01)00884838099210(21)165 183 (01)00884838099210(21)183 189 (01)00884838099210(21)189 182 (01)00884838099210(21)182 185 (01)00884838099210(21)185 159 (01)00884838099210(21)159 184 (01)00884838099210(21)184 190 (01)00884838099210(21)190 174 (01)00884838099210(21)174 177 (01)00884838099210(21)177 (3)Model: 722280 20 (01)00884838103276(21)20. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Distribuidor
- Worldwide distribution - US Nationwide distribution in the states of NY, PA, PR, TN, TX, VA, WA and the countries of Algeria, Armenia, Austria, Brazil, China, Colombia, Croatia, Czech Republic, Ecuador, France, Germany, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Korea, Republic of Kuwait, Lithuania, Macao, Macedonia, Malaysia, Morocco, Netherlands, Norway, Oman, Palestine, State of Panama, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, Tunisia, Turkey, Ukraine, United Kingdom.
- Lote
- OUS: (1) Model: 722221 SN: 66 (01)00884838099203(21)66; SN: 91 (01)00884838099203(21)91; SN: 104 (01)00884838099258(21)104, (2) Model: 722222 SN: UDI: 165 (01)00884838099210(21)165 183 (01)00884838099210(21)183 189 (01)00884838099210(21)189 182 (01)00884838099210(21)182 185 (01)00884838099210(21)185 159 (01)00884838099210(21)159 184 (01)00884838099210(21)184 190 (01)00884838099210(21)190 174 (01)00884838099210(21)174 177 (01)00884838099210(21)177 (3)Model: 722280 20 (01)00884838103276(21)20
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Netherlands
Retiros relacionados
- FDAPhilips Azurion 5 M12. Model Number: 722227.4 de noviembre de 2026
- FDAPhilips Azurion 3 M12. Model Numbers: 722063, 722221.4 de noviembre de 2026
- FDAPhilips Azurion 7 M20. Model Numbers: 722079, 722224, 722282.4 de noviembre de 2026
- FDAPhilips Azurion 5 M20. Model Numbers: 722228, 722281.4 de noviembre de 2026
- FDAAutomated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella28 de febrero de 2024
- FDAAzurion 5 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally28 de febrero de 2024
Preguntas sobre este retiro
¿Azurion 3 with a Certeray generator -To perform image guidance in diagnostic, interventional, and minimally invasive surgery procedures Models: (1) 722221 (2) 722222 (3) 722280 fue retirado del mercado?
Sí. FDA publicó el retiro Z-1176-2024 el 28 de febrero de 2024. Peligro reportado: Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter