Arthrex Shoulder Suspension Tower; Catalog Number: AR-1650-01 Product Usage: The Arthrex Lateral Decubitus Shoulder Positioner is intended to allow for the controlled manipulation of the arm and shoulder joint during arthroscopic surgical procedures.
Publicado el 22 de febrero de 2017 · Inicio del retiro: 3 de enero de 2017
Resumen
FDA registra el retiro Z-1198-2017 de dispositivos médicos de Clase II, publicado el 22 de febrero de 2017 e iniciado el 3 de enero de 2017. Empresa responsable: Allen Medical Systems, Inc.; Estado: Finalizado; Cantidad reportada: 112 devices; Lote: Serial Numbers: 501798, 501799, 501800, 501801, 501802, 501803, 501804, 501805, 501806, 501807, 501808, 501809, 501810, 501811, 501812, 501813, 501814, 501815, 501816, 501817, 501818, 501819, 501820, 501821, 501822, 501823, 501824, 501825, 501826, 501827, 501828, 501829, 501830, 501831, 501832, 501833, 501834, 501835, 501836, 501837, 501838, 501839, 501840, 501841, 501842, 501848, 501849, 501850, 501851, 501852, 501853, 501854, 501855, 501856, 501857, 501863, 501864, 501865, 501866, 501867, 501868, 501869, 501870, 501871, 501872, 501878, 501879, 501880, 501881, 501882, 501883, 501884, 501885, 501886, 501887, 501888, 501889, 501890, 501891, 501892, 501893, 501894, 501895, 501896, 501897, 510733, 510734, 510735, 510736, 510737, 510738, 510739, 510740, 510741, 510742, 510743, 510744, 510745, 510746, 510747, 510748, 510749, 510750, 510751, 510752, 510834, 510835, 510836, 510837, 510838, 510839, 510840. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
During use, there is a potential scenario which could allow the fastening screws attaching covers to come loose and potentially land in the sterile field causing a delay in surgery.
Qué hacer
Voluntary: Firm initiated · E-Mail
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Allen Medical Systems, Inc.
- Distribuidor
- US state of FL
- Lote
- Serial Numbers: 501798, 501799, 501800, 501801, 501802, 501803, 501804, 501805, 501806, 501807, 501808, 501809, 501810, 501811, 501812, 501813, 501814, 501815, 501816, 501817, 501818, 501819, 501820, 501821, 501822, 501823, 501824, 501825, 501826, 501827, 501828, 501829, 501830, 501831, 501832, 501833, 501834, 501835, 501836, 501837, 501838, 501839, 501840, 501841, 501842, 501848, 501849, 501850, 501851, 501852, 501853, 501854, 501855, 501856, 501857, 501863, 501864, 501865, 501866, 501867, 501868, 501869, 501870, 501871, 501872, 501878, 501879, 501880, 501881, 501882, 501883, 501884, 501885, 501886, 501887, 501888, 501889, 501890, 501891, 501892, 501893, 501894, 501895, 501896, 501897, 510733, 510734, 510735, 510736, 510737, 510738, 510739, 510740, 510741, 510742, 510743, 510744, 510745, 510746, 510747, 510748, 510749, 510750, 510751, 510752, 510834, 510835, 510836, 510837, 510838, 510839, 510840
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDASTERRAD NX Sterilization System, Product Code: 10033, 10033-00222 de febrero de 2017
- FDA6-hole Pubic Symphysis plate (PN 70-0451). The pelvic bone plates for the pelvic ring are a series of plates22 de febrero de 2017
- FDAHARVEST NEEDLE 15GA X 2 Intended for the purpose of harvesting bone and/or bone marrow specimens22 de febrero de 2017
- FDAStarClose SE Vascular Closure System, Part No. 14679-01, 14679-02. The UDI is 08717648079467. The GMDN is22 de febrero de 2017
- FDAArthex S3 Arm Sleeve Connector; Catalog Number AR-1650-02; manufactured and distributed between March and4 de enero de 2017
- FDAAllen Traction Boot, Modular, OT 1000 Traction Device Boot Assembly, Fracture Table Orthopedic Leg Holders,27 de marzo de 2013
Preguntas sobre este retiro
¿Arthrex Shoulder Suspension Tower; Catalog Number: AR-1650-01 Product Usage: The Arthrex Lateral Decubitus Shoulder Positioner is intended to allow for the controlled manipulation of the arm and shoulder joint during arthroscopic surgical procedures. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1198-2017 el 22 de febrero de 2017. Peligro reportado: During use, there is a potential scenario which could allow the fastening screws attaching covers to come loose and potentially land in the sterile field causing a delay in surgery.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · E-Mail