REF 623411 L, S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Left, Small, Sterile R Qty 1 The S-ROM Hinged knee femoral component is used in revision surgeries when there is significant bone loss or ligament instability and additional support is needed. S-Rom femorals and inserts are brought into nearly every revision case as backup should the MCL become compromised during the procedure. If additional constraint is required the surgeon can use the S-ROM femoral and insert. S-ROM femorals are used in less than 10% of knee revision procedures, but are present almost all the time.
Publicado el 19 de marzo de 2014 · Inicio del retiro: 24 de febrero de 2014
Resumen
FDA registra el retiro Z-1205-2014 de dispositivos médicos de Clase II, publicado el 19 de marzo de 2014 e iniciado el 24 de febrero de 2014. Empresa responsable: DePuy Orthopaedics, Inc.; Estado: Finalizado; Cantidad reportada: 134; Lote: E1VHS4, 430482, 338164, 373256, 395802, 409578, 448220, 409582, 238688, 430483, 306508, 329330, 264451, 448222, 216275, 373255, 346439, 409580, 338165, 250929, 360638, D65FJ4, 216277, 395804, EJ7F14, 254740, 302815, FF8DW4, 314821, 464642, 338166, FA4FP4, FA4FT4, 238411LC, 464643, 278448, 409579, 464640, 395803, 088411LA, 308248, FF8DX4, 196411LA, 448227, C1VCE4, 402352, 346437, 366433, 238687, 172188, 329332, 250931, 464641, 216274, 302813, 295384, 264453, 337411LA, 120084, 236445, 289043, 329331, 308247, and 395806. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
DePuy Orthopaedics has identified the potential for holes to develop in the inner and outer flexible pouches that form the sterile barrier for S-ROM® Noiles Rotating Hinge Femur with Pin devices. The outer carton and shrink wrap are intact. A package redesign is underway to resolve this issue. S-ROM Noiles Rotating Hinge Femur with Pin devices will remain in US Distributors inventory during the p
Qué hacer
Voluntary: Firm initiated · E-Mail
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- DePuy Orthopaedics, Inc.
- Distribuidor
- Worldwide Distribution - USA (nationwide) and Internationally to Canada, Australia, Austria, China, Czech Republic, Denmark, Dubai / UAE, Estonia, France, Germany, Hungary, Iberia, India, Ireland, Israel, Italy, Malaysia, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Russia, Singapore, Slovenia, Spain, South Africa, Sweden, Switzerland, Turkey, United Kingdom, and Vietnam.
- Lote
- E1VHS4, 430482, 338164, 373256, 395802, 409578, 448220, 409582, 238688, 430483, 306508, 329330, 264451, 448222, 216275, 373255, 346439, 409580, 338165, 250929, 360638, D65FJ4, 216277, 395804, EJ7F14, 254740, 302815, FF8DW4, 314821, 464642, 338166, FA4FP4, FA4FT4, 238411LC, 464643, 278448, 409579, 464640, 395803, 088411LA, 308248, FF8DX4, 196411LA, 448227, C1VCE4, 402352, 346437, 366433, 238687, 172188, 329332, 250931, 464641, 216274, 302813, 295384, 264453, 337411LA, 120084, 236445, 289043, 329331, 308247, and 395806
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAATTUNE AFFIXIUM Cementless Fixed Bearing Knee with 3DP Technology, Size 6. Component of the ATTUNE Total Knee10 de abril de 2024
- FDAAttune Revision Limb Preservation System (LPS) Tibial Insert XSM 12MM, Part Number 15176021229 de marzo de 2023
- FDAAttune Revision Limb Preservation System (LPS) Tibial Insert SM 12MM, Part Number 15176031229 de marzo de 2023
- FDAATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation27 de julio de 2022
- FDAREF 623401 R, S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Right, Medium, Sterile19 de marzo de 2014
- FDAREF 623411 R S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Right, Small, Sterile R19 de marzo de 2014
Preguntas sobre este retiro
¿REF 623411 L, S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Left, Small, Sterile R Qty 1 The S-ROM Hinged knee femoral component is used in revision surgeries when there is significant bone loss or ligament instability and additional support is needed. S-Rom femorals and inserts are brought into nearly every revision case as backup should the MCL become compromised during the procedure. If additional constraint is required the surgeon can use the S-ROM femoral and insert. S-ROM femorals are used in less than 10% of knee revision procedures, but are present almost all the time. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1205-2014 el 19 de marzo de 2014. Peligro reportado: DePuy Orthopaedics has identified the potential for holes to develop in the inner and outer flexible pouches that form the sterile barrier for S-ROM® Noiles Rotating Hinge Femur with Pin devices. The outer carton and shrink wrap are intact. A package redesign is underway to resolve this issue. S-ROM Noiles Rotating Hinge Femur with Pin devices will remain in US Distributors inventory during the p
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · E-Mail