Philips Model EPIQ 7 Ultrasound System. Intended for diagnostic ultrasound imaging and fluid flow analysis. Catalogue numbers 795200 / 795201 / 795202 / 795204 / 795205 / 795206 EPIQ is intended for diagnostic ultrasound imaging and fluid flow analysis. The collection of Doppler velocity results values are used to assist in diagnosing the existence, location and severity of cardiac or vasculare stenosis.
Publicado el 26 de marzo de 2014 · Inicio del retiro: 28 de febrero de 2014
Resumen
FDA registra el retiro Z-1264-2014 de dispositivos médicos de Clase II, publicado el 26 de marzo de 2014 e iniciado el 28 de febrero de 2014. Empresa responsable: Philips Medical Systems North America Inc.; Estado: Finalizado; Cantidad reportada: 1267 total :670 units in US and 1267 units Internationally (which includes 11 units in Canada); Lote: All systems below software v1.1.2 Which includes software v1 and v 1.1 (All lot codes shipped prior to 2/28/2014). La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
A problem has been detected with Philips Model EPIQ 5 and 7 Ultrasound System. The system will report incorrect Doppler velocities if you use High Pulse Repetition Frequency (HPRF) with multiple sample volume gates. While in High Pulse Repetition Frequency (HPRF) in Display Zoom Velocity Scale using manual spectral Doppler trace, the velocities in the velocity results table may be overstated b
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Philips Medical Systems North America Inc.
- Distribuidor
- Worldwide Distribution: US (nationwide) including states of: AL, AK, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NM, NJ, NM, NY, OH, OK, PA, SC, TN, TX, VA, WA, and WV; and countries of: Australia, Austria, Belgium, Canada, Chili, Ecuador, Egypt, Finland, France, Germany, Greece, India, Ireland, Italy, Japan, Korea Rep, Malaysia, Netherlands, New Zealand, Norway, Peru, Philippines, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom.
- Lote
- All systems below software v1.1.2 Which includes software v1 and v 1.1 (All lot codes shipped prior to 2/28/2014).
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAPhilips HeartStart FR3 Automated External Defibrillator (AED), models 861388 and 861389. The HeartStart FR3 is26 de marzo de 2014
- FDAPhilips Model EPIQ 5 Ultrasound System. Intended for diagnostic ultrasound imaging and fluid flow analysis.26 de marzo de 2014
- FDAPhilips HeartStart Home and OnSite (HS1) automated external defibrillators (AEDs). Product Usage: The models26 de marzo de 2014
- FDAHalogen Lamp component in RetCam 3 System - Clarity Medical Systems, Inc; Pleasanton, CA. General ophthalmic26 de marzo de 2014
- FDANatural-Knee® II System: POLY ULTRA TIB LT SZ00/9M POLY ULTRA TIB LT SZ0/9MM POLY ULTRA TIB LT SZ0/19M POLY26 de marzo de 2014
- FDAEPIQ 7 Ultrasound System, Model 795200, 795201, 795202. Intended for diagnostic ultrasound imaging and fluid27 de noviembre de 2013
Preguntas sobre este retiro
¿Philips Model EPIQ 7 Ultrasound System. Intended for diagnostic ultrasound imaging and fluid flow analysis. Catalogue numbers 795200 / 795201 / 795202 / 795204 / 795205 / 795206 EPIQ is intended for diagnostic ultrasound imaging and fluid flow analysis. The collection of Doppler velocity results values are used to assist in diagnosing the existence, location and severity of cardiac or vasculare stenosis. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1264-2014 el 26 de marzo de 2014. Peligro reportado: A problem has been detected with Philips Model EPIQ 5 and 7 Ultrasound System. The system will report incorrect Doppler velocities if you use High Pulse Repetition Frequency (HPRF) with multiple sample volume gates. While in High Pulse Repetition Frequency (HPRF) in Display Zoom Velocity Scale using manual spectral Doppler trace, the velocities in the velocity results table may be overstated b
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter