Philips HeartStart FR3 Automated External Defibrillator (AED), models 861388 and 861389. The HeartStart FR3 is a battery powered automated external defibrillator (AED) designed for use by responders who have been trained in Basic Life Support (BLS), Advanced Life Support (ALS), or another physician-authorized emergency medical response program. The HeartStart FR3 is used to treat suspected victims of ventricular fibrillation (VF), the most common cause of sudden cardiac arrest (SCA), and certain ventricular tachycardia (VTs). Both models are used with disposable defibrillator pads applied to potential victims of SCA with the following symptoms: -Unresponsiveness -Absence of normal breathing If in doubt, apply the pads. The HeartStart FR3 is intended for adults and children over 55 pounds (25 kg) or 8 years old. Both models 861388 and 861389 are also intended for children under 55 pounds (25 kg) or 8 years old when used with the optional Infant/Child Key. If the Infant/Child Key is not available, or you are uncertain of the child's age or weight, do not delay treatment.

Publicado el 26 de marzo de 2014 · Inicio del retiro: 6 de febrero de 2014

Resumen

FDA registra el retiro Z-1225-2014 de dispositivos médicos de Clase II, publicado el 26 de marzo de 2014 e iniciado el 6 de febrero de 2014. Empresa responsable: Philips Medical Systems North America Inc.; Estado: Finalizado; Cantidad reportada: 251 units total (128 in USA and 123 untils outside USA); Lote: Serial numbers of units distributed in the USA: C13A-00592, C13C-00131, C13C-00150, C13C-00270, C13C-00679, C13C-00681, C13C-00769, C13D-00048, C13D-00197, C13E-00088, C13E-00168, C13E-00234, C13E-00242, C13E-00245, C13E-00326, C13E-00340, C13E-00368, C13E-00372, C13E-00380, C13E-00382, C13E-00390, C13E-00406, C13E-00495, C13E-00507, C13E-00511, C13E-00516, C13E-00526, C13E-00537, C13F-00011, C13F-00012, C13F-00013, C13F-00022, C13F-00024, C13F-00027, C13F-00031, C13F-00032, C13F-00033, C13F-00035, C13F-00038, C13F-00043, C13F-00047, C13F-00053, C13F-00054, C13F-00059, C13F-00071, C13F-00077, C13F-00078, C13F-00102, C13F-00104, C13F-00110, C13F-00114, C13F-00118, C13F-00126, C13F-00131, C13F-00137, C13F-00138, C13F-00140, C13F-00143, C13F-00146, C13F-00148, C13F-00168, C13F-00172, C13F-00173, C13F-00174, C13F-00180, C13F-00181, C13F-00198, C13F-00202, C13F-00212, C13F-00214, C13F-00215, C13F-00217, C13F-00218, C13F-00224, C13F-00225, C13F-00227, C13F-00232, C13F-00233, C13F-00236, C13F-00238, C13F-00239, C13F-00240, C13F-00244, C13F-00246, C13F-00250, C13F-00255, C13F-00266, C13F-00273, C13F-00274, C13F-00275, C13F-00278, C13F-00284, C13F-00291, C13F-00294, C13F-00307, C13F-00309, C13F-00318, C13F-00321, C13F-00322, C13F-00323, C13F-00329, C13F-00330, C13F-00331, C13F-00332, C13F-00336, C13F-00348, C13F-00353, C13F-00468, C13F-00543, C13F-00559, C13F-00567, C13F-00578, C13F-00590, C13F-00593, C13F-00595, C13F-00599, C13F-00604, C13F-00607, C13F-00651, C13F-00778, C13G-00008, C13G-00010, C13G-00035, C13G-00056, C13G-00057, C13G-00282, C13G-00284, and C13H-00062. Serial numbers of units outside the USA: C13C-00004, C13C-00256, C13C-00305, C13C-00323, C13C-00608, C13C-00639, C13E-00009, C13E-00021, C13E-00036, C13E-00048, C13E-00057, C13E-00058, C13E-00065, C13E-00163, C13E-00250, C13E-00265, C13E-00273, C13E-00311, C13E-00323, C13E-00328, C13E-00339, C13E-00341, C13E-00345, C13E-00346, C13E-00349, C13E-00350, C13E-00351, C13E-00358, C13E-00393, C13E-00395, C13E-00404, C13E-00407, C13E-00409, C13E-00424, C13E-00437, C13E-00441, C13E-00452, C13E-00454, C13E-00455, C13E-00499, C13E-00503, C13E-00518, C13E-00527, C13E-00535, C13E-00548, C13F-00010, C13F-00280, C13F-00286, C13F-00293, C13F-00298, C13F-00302, C13F-00303, C13F-00305, C13F-00306, C13F-00316, C13F-00317, C13F-00343, C13F-00345, C13F-00346, C13F-00358, C13F-00363, C13F-00369, C13F-00371, C13F-00381, C13F-00385, C13F-00391, C13F-00400, C13F-00420, C13F-00424, C13F-00428, C13F-00436, C13F-00473, C13F-00498, C13F-00539, C13F-00565, C13F-00579, C13F-00580, C13F-00583, C13F-00586, C13F-00587, C13F-00606, C13F-00619, C13F-00627, C13F-00629, C13F-00649, C13F-00666, C13F-00668, C13F-00677, C13F-00678, C13F-00682, C13F-00685, C13F-00687, C13F-00726, C13F-00742, C13F-00746, C13F-00748, C13F-00755, C13F-00758, C13F-00760, C13F-00782, C13F-00797, C13F-00815, C13F-00870, C13F-00889, C13F-00896, C13F-00938, C13G-00026, C13G-00039, C13G-00084, C13G-00126, C13G-00132, C13G-00142, C13G-00237, C13G-00247, C13G-00293, C13G-00315, C13G-00331, C13G-00337, C13G-00345, C13G-00407, C13G-00435, C13G-00451, and C13H-00035. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Philip's is recalling HeartStart FR3 Automated External Defibrillator (AED) because the Printed Circuit Board (PCB), which contains the circuitry that supplies energy to the device, could become separated from the electrical contact points that hold the board in place. In the event the PCB becomes separated from the electrical contact points, the AED may not be ready to deliver a shock in the event of an emergency.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Philips Medical Systems North America Inc.
Distribuidor
Worldwide Distribution - US Distribution including the states of AL, CA, CO, WA DC, FL, GA, HI, KS, MD, ME, MI, MS, NC, NM, NY, PA, TX, VA, WA and WI., and the countries of Australia, Austria, Canada, France, Germany, India, Italy, Ireland, Japan, Malaysia, Netherlands, Norway, Switzerland, Taiwan and United Kingdom.
Lote
Serial numbers of units distributed in the USA: C13A-00592, C13C-00131, C13C-00150, C13C-00270, C13C-00679, C13C-00681, C13C-00769, C13D-00048, C13D-00197, C13E-00088, C13E-00168, C13E-00234, C13E-00242, C13E-00245, C13E-00326, C13E-00340, C13E-00368, C13E-00372, C13E-00380, C13E-00382, C13E-00390, C13E-00406, C13E-00495, C13E-00507, C13E-00511, C13E-00516, C13E-00526, C13E-00537, C13F-00011, C13F-00012, C13F-00013, C13F-00022, C13F-00024, C13F-00027, C13F-00031, C13F-00032, C13F-00033, C13F-00035, C13F-00038, C13F-00043, C13F-00047, C13F-00053, C13F-00054, C13F-00059, C13F-00071, C13F-00077, C13F-00078, C13F-00102, C13F-00104, C13F-00110, C13F-00114, C13F-00118, C13F-00126, C13F-00131, C13F-00137, C13F-00138, C13F-00140, C13F-00143, C13F-00146, C13F-00148, C13F-00168, C13F-00172, C13F-00173, C13F-00174, C13F-00180, C13F-00181, C13F-00198, C13F-00202, C13F-00212, C13F-00214, C13F-00215, C13F-00217, C13F-00218, C13F-00224, C13F-00225, C13F-00227, C13F-00232, C13F-00233, C13F-00236, C13F-00238, C13F-00239, C13F-00240, C13F-00244, C13F-00246, C13F-00250, C13F-00255, C13F-00266, C13F-00273, C13F-00274, C13F-00275, C13F-00278, C13F-00284, C13F-00291, C13F-00294, C13F-00307, C13F-00309, C13F-00318, C13F-00321, C13F-00322, C13F-00323, C13F-00329, C13F-00330, C13F-00331, C13F-00332, C13F-00336, C13F-00348, C13F-00353, C13F-00468, C13F-00543, C13F-00559, C13F-00567, C13F-00578, C13F-00590, C13F-00593, C13F-00595, C13F-00599, C13F-00604, C13F-00607, C13F-00651, C13F-00778, C13G-00008, C13G-00010, C13G-00035, C13G-00056, C13G-00057, C13G-00282, C13G-00284, and C13H-00062. Serial numbers of units outside the USA: C13C-00004, C13C-00256, C13C-00305, C13C-00323, C13C-00608, C13C-00639, C13E-00009, C13E-00021, C13E-00036, C13E-00048, C13E-00057, C13E-00058, C13E-00065, C13E-00163, C13E-00250, C13E-00265, C13E-00273, C13E-00311, C13E-00323, C13E-00328, C13E-00339, C13E-00341, C13E-00345, C13E-00346, C13E-00349, C13E-00350, C13E-00351, C13E-00358, C13E-00393, C13E-00395, C13E-00404, C13E-00407, C13E-00409, C13E-00424, C13E-00437, C13E-00441, C13E-00452, C13E-00454, C13E-00455, C13E-00499, C13E-00503, C13E-00518, C13E-00527, C13E-00535, C13E-00548, C13F-00010, C13F-00280, C13F-00286, C13F-00293, C13F-00298, C13F-00302, C13F-00303, C13F-00305, C13F-00306, C13F-00316, C13F-00317, C13F-00343, C13F-00345, C13F-00346, C13F-00358, C13F-00363, C13F-00369, C13F-00371, C13F-00381, C13F-00385, C13F-00391, C13F-00400, C13F-00420, C13F-00424, C13F-00428, C13F-00436, C13F-00473, C13F-00498, C13F-00539, C13F-00565, C13F-00579, C13F-00580, C13F-00583, C13F-00586, C13F-00587, C13F-00606, C13F-00619, C13F-00627, C13F-00629, C13F-00649, C13F-00666, C13F-00668, C13F-00677, C13F-00678, C13F-00682, C13F-00685, C13F-00687, C13F-00726, C13F-00742, C13F-00746, C13F-00748, C13F-00755, C13F-00758, C13F-00760, C13F-00782, C13F-00797, C13F-00815, C13F-00870, C13F-00889, C13F-00896, C13F-00938, C13G-00026, C13G-00039, C13G-00084, C13G-00126, C13G-00132, C13G-00142, C13G-00237, C13G-00247, C13G-00293, C13G-00315, C13G-00331, C13G-00337, C13G-00345, C13G-00407, C13G-00435, C13G-00451, and C13H-00035.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Philips HeartStart FR3 Automated External Defibrillator (AED), models 861388 and 861389. The HeartStart FR3 is a battery powered automated external defibrillator (AED) designed for use by responders who have been trained in Basic Life Support (BLS), Advanced Life Support (ALS), or another physician-authorized emergency medical response program. The HeartStart FR3 is used to treat suspected victims of ventricular fibrillation (VF), the most common cause of sudden cardiac arrest (SCA), and certain ventricular tachycardia (VTs). Both models are used with disposable defibrillator pads applied to potential victims of SCA with the following symptoms: -Unresponsiveness -Absence of normal breathing If in doubt, apply the pads. The HeartStart FR3 is intended for adults and children over 55 pounds (25 kg) or 8 years old. Both models 861388 and 861389 are also intended for children under 55 pounds (25 kg) or 8 years old when used with the optional Infant/Child Key. If the Infant/Child Key is not available, or you are uncertain of the child's age or weight, do not delay treatment. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1225-2014 el 26 de marzo de 2014. Peligro reportado: Philip's is recalling HeartStart FR3 Automated External Defibrillator (AED) because the Printed Circuit Board (PCB), which contains the circuitry that supplies energy to the device, could become separated from the electrical contact points that hold the board in place. In the event the PCB becomes separated from the electrical contact points, the AED may not be ready to deliver a shock in the event of an emergency.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter