A) Product Code 2N8371: Non-DEHP Y-Type Catheter Extension Set Kit, 6.0", Vol. 1.0 mL, 2 Luer Activated Valves, Male Luer Lock Adapter, Single use only Product Usage: The IV extension tubing kits are intended to provide users with a ready to use set up for the administration and withdrawal of fluids avoiding the need to acquire/assemble components from various manufacturers. The IV extension tubing sets are indicated for use with a vascular access device for administration and withdrawal of fluids.
Publicado el 15 de mayo de 2013 · Inicio del retiro: 28 de marzo de 2013
Resumen
FDA registra el retiro Z-1271-2013 de dispositivos médicos de Clase II, publicado el 15 de mayo de 2013 e iniciado el 28 de marzo de 2013. Empresa responsable: Baxter Healthcare Corp.; Estado: Finalizado; Cantidad reportada: A) Product Code 2N8371: 513,000 units; Lote: A) Product Code 2N8371: Lots R12E03108, R12H27085, R12H28075, R12H29057, R12H30139, R12103092, R12104058, R12105022, R12106061, R12110113, R12119056, R12120112, R12121151, R12122050, R12124106, R12124130, R12J25045, R12K03057, R12K19046, R12K21034, R12K21083, R12L05118, R12L08054, R12L18111, R12Ll9101, R13A08063, R13A08147, Rl3A09137, R13A10101, R13A14129, R13A15050. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Customers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile barrier properties.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Baxter Healthcare Corp.
- Distribuidor
- Worldwide Distribution - USA Nationwide and the countries of Puerto Rico, Brazil, Canada and Columbia.
- Lote
- A) Product Code 2N8371: Lots R12E03108, R12H27085, R12H28075, R12H29057, R12H30139, R12103092, R12104058, R12105022, R12106061, R12110113, R12119056, R12120112, R12121151, R12122050, R12124106, R12124130, R12J25045, R12K03057, R12K19046, R12K21034, R12K21083, R12L05118, R12L08054, R12L18111, R12Ll9101, R13A08063, R13A08147, Rl3A09137, R13A10101, R13A14129, R13A15050
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAGAMBRO Capillary Dialyzer, Polyflux(R) Revaclear MAX, STERILE, Product code 110634 Product Usage: The16 de noviembre de 2016
- FDASynovis VASCU-GUARD Peripheral Vascular Patch in the following sizes and product codes: 1 cm x 6 cm, product24 de agosto de 2016
- FDA0.9% Sodium Chloride Injection, USP, 250 mL VIAFLEX Plastic Container Bag, Rx Only, Manufactured by Baxter15 de junio de 2016
- FDADextrose Injection,USP,70% 2000 mL VIAFLEX Plastic Container Bag, Pharmacy Bulk Package, Not For Direct15 de junio de 2016
- FDAA) Product Code 2N8220: Non-DEHP Catheter Extension Set Kit, Single use only, includes: 1) 7.7", Vol. 1.10 mL,15 de mayo de 2013
- FDASiemens Healthcare Diagnostics Inc. MIcroScan MICroSTREP plus Type 1 B1027-201 panels. .MICroSTREP panels are15 de mayo de 2013
Preguntas sobre este retiro
¿A) Product Code 2N8371: Non-DEHP Y-Type Catheter Extension Set Kit, 6.0", Vol. 1.0 mL, 2 Luer Activated Valves, Male Luer Lock Adapter, Single use only Product Usage: The IV extension tubing kits are intended to provide users with a ready to use set up for the administration and withdrawal of fluids avoiding the need to acquire/assemble components from various manufacturers. The IV extension tubing sets are indicated for use with a vascular access device for administration and withdrawal of fluids. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1271-2013 el 15 de mayo de 2013. Peligro reportado: Customers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile barrier properties.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter