FDAFDA · Alimentos y medicamentos·N.º oficial Z-1399-2025
VERICIS, Merge Cardio, Model/Catalog Number: Version 10.x, Software Version: 10.0, 10.1.1.1, 10.1.2, 10.2.0, 10.2 P2 , 10.3; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
Publicado el 26 de marzo de 2025 · Inicio del retiro: 19 de febrero de 2025
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
Qué hacer
Voluntary: Firm initiated · E-Mail
Detalles del retiro
- Fabricante
- Merge Healthcare, Inc.
- Distribuidor
- US
- Lote
- Software Versions and Part Numbers: 10.0: Part Numbers: 88-00527-00, 88-00528-00, 88-00529-00, 88-00530-00, 88-00531-00, 88-00532-00, 88-00533-00, 70-00008-00; 10.1.1.1: Part Numbers: 88-01094-00; 10.1.2: Part Numbers: 88-01074-00, 88-01075-00, 88-01076-00, 88-01077-00, 88-01078-00, 88-01079-00, 88-00781-00, 88-00803-00, 88-01080-00, 70-00473-00; 10.2.0: Part Numbers: 88-01115-00, 88-01116-00, 88-01117-00, 88-01118-00, 88-01119-00, 88-01120-00, 88-01121-00, 88-01122-00, 88-01123-00, 70-00491-00, 10.2 P2: Part Numbers: 88-01203-00 10.3 P1 UDI - (01)00842000100041(10)10.3.0.2009(11)170309
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?