Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1404-2020

MicroVention Terumo Microplex 10, Platinum Coil System, Endovascular Embolization Coil, Compass 10, Sterile, Rx, REF numbers 100203CMSR-V, 100408CMSR-V, 100412CMSR-V, 100510CMSR-V, 100516CMSR-V, 100612CMSR-V, 100618CMSR-V, and 100721CMSR-V. Product Usage: The MCS is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. The firm name on the label is MicroVention, Inc., Tustin, CA.

Publicado el 4 de marzo de 2020 · Inicio del retiro: 22 de noviembre de 2019

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The devices may be missing the implant coil.

Qué hacer

Voluntary: Firm initiated · E-Mail

Detalles del retiro

Fabricante
Microvention, Inc.
Distribuidor
All distribution was OUS. The countries receiving affected product are: Albania, Austria, Belarus, Belgium, Brazil, Bulgaria, China, Cyprus, Czech Republic, Denmark, Egypt, Finalnd, France, Georgia, Germany, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Malta, Morocco, Netherlands, Norway, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Slovak Republic, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
Lote
Catalog/lot numbers: 100203CMSR-V - 1903085WP; 100408CMSR-V - 1903085WP and 1904015PP; 100412CMSR-V - 1904225RP, 1905205QP, and 1904035PP; 100510CMSR-V - 1905205QP and 1904085WP; 100516CMSR-V - 1903135WP and 1905155WP; 100612CMSR-V - 1903225WP; 100618CMSR-V - 1903135QP and 1904085WP; and 100721CMSR-V - 1903085RP.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States