FDAFDA · Alimentos y medicamentos·N.º oficial Z-1420-2013
REF 8229705 x1 NIM TriVantage EMG Endotracheal Tube 5.0 mm I.D. x 6.5 mm O.D. Medtronic Xomed, Inc., 6743 Southpoint Drive North, Jacksonville, FL 32216-0980 USA STERILE Rx Only The electrodes are designed to make contact with the patient's vocal cords to facilitate electromyographic (EMG) monitoring of the laryngeal musculature during surgery when connected to a multi-channel EMG neuromonitoring device.
Publicado el 12 de junio de 2013 · Inicio del retiro: 4 de marzo de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
In March 2013, Medtronic issued a recall of the NIM TriVantage EMG Endotracheal Tube due to cuff leakage in the intraoperative cuff deflation which then requires re-inflation or replacement of the deflated tube in order to ensure the continued ventilation of the patient.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Medtronic Xomed, Inc.
- Distribuidor
- Worldwide distribution: AL, AR, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, MA, MD, MI, MN, MT, NC, NH, NJ, NY, OH, PA, RI, SC, TX, UT, VA, and WA; and countries of : Australia, Canada, Germany, Italy, Norway, Poland and United Kingdom.
- Lote
- Product Catalog Numbers: REF 8229705, 8229706, 8229707, 8229708, 8229709, 8229735, 8229736, 8229737, 8229738, 8229739. Lot Numbers: 205830052 to 206486732. 0206516104, 0206516105, 0206516106, 0206516108, 0206520224, 0206520225, 0206520226, 0206520227, 0206520228, 0206520358, 0206542163, 0206545356, 0206545502.
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States