FDAFDA · Alimentos y medicamentos·N.º oficial Z-1462-2014

The OEC 9900 Elite mobile fluoroscopy system (image intensified fluoroscopic x-ray) is designed to provide fluoroscopic and spot-film images of the patient during diagnostic surgical and interventional procedure. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedure. The system may be used for other imaging applications at the physician's discretion.

Publicado el 23 de abril de 2014 · Inicio del retiro: 31 de marzo de 2014

Resumen

FDA registra el retiro Z-1462-2014 de dispositivos médicos de Clase II, publicado el 23 de abril de 2014 e iniciado el 31 de marzo de 2014. Empresa responsable: GE OEC Medical Systems, Inc; Estado: Finalizado; Cantidad reportada: 453 units; Lote: Model Numbers - OEC 9900 Elite. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

GE Healthcare has identified a potential safety issue related to the screws that attach the workstation monitor spring arm to the extension arm on certain OEC 9900 systems, causing the spring arm and monitor to fall forward.

Qué hacer

Voluntary: Firm initiated · Telephone

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
GE OEC Medical Systems, Inc
Distribuidor
Worldwide Distribution - USA, and the countries of Australia, Argentina, Canada, China, Cambodia, Columbia, Japan, Ecuador, Japan, Korea, Hong Kong, India, Israel, Libya, Mexico, Nigeria, Peru, Russia, Thailand, Taiwan, Saudi Arabia and Singapore.
Lote
Model Numbers - OEC 9900 Elite
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿The OEC 9900 Elite mobile fluoroscopy system (image intensified fluoroscopic x-ray) is designed to provide fluoroscopic and spot-film images of the patient during diagnostic surgical and interventional procedure. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedure. The system may be used for other imaging applications at the physician's discretion. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1462-2014 el 23 de abril de 2014. Peligro reportado: GE Healthcare has identified a potential safety issue related to the screws that attach the workstation monitor spring arm to the extension arm on certain OEC 9900 systems, causing the spring arm and monitor to fall forward.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Telephone