Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1464-2026

Medline Kits containing Tego Connectors Medline DIALYSIS ON/OFF KIT SKU EBSI1746

Publicado el 4 de marzo de 2026 · Inicio del retiro: 8 de enero de 2026

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Medline Industries, LP
Distribuidor
US Nationwide distribution.
Lote
UDI-DI (ea) 10653160373582 UDI-DI (case) 00653160373585 Kit lots 2024051380 2024052080 2024052180 2024052280 2024071580 2025061790 2025072990
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta