VITROS Chemistry Products Li (Lithium) Slides, REF/Catalog Number/Product Code 1632660 (18 slide format, 90 slides per sales unit) and REF/Catalog Number/Product Code 8318925 (60 slide format / 300 slides per sales unit), Unique Identifier Nos. 10758750009466 and 10758750004720, RX ONLY, IVD --Ortho Clinical Diagnostics, Inc. Rochester, NY 14626 --- Background: The VITROS Li Slide is a multilayered, analytical element coated on a polyester support. During the coating process, thin layers of reagents are coated onto a polyester support layer to create a spreading layer and a chemical (reagent) layer. During the coating process rolls of materials are produced. These rolls are segmented into slits. There are several slits in each roll. Each slit is then cut into small chemistry chips that are positioned between plastic to create the final Microslide format. Eighteen (18) or sixty (60) slides are placed into each plastic cartridge for use. Each slide is a device.
Publicado el 22 de marzo de 2017 · Inicio del retiro: 1 de febrero de 2017
Resumen
FDA registra el retiro Z-1471-2017 de dispositivos médicos de Clase II, publicado el 22 de marzo de 2017 e iniciado el 1 de febrero de 2017. Empresa responsable: Ortho-Clinical Diagnostics; Estado: Finalizado; Cantidad reportada: US: 94 sales units; OUS: 136 sales units; Lote: REF/Product Code 1632660: Lot 3124-0108-6568, exp. 2018- 04-01; REF/Product Code 8318925: Lot 3124-0108-6569, exp. 2018- 05-01. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Intermittent imprecision in results generated using two specific lots of VITROS Li Slides.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Ortho-Clinical Diagnostics
- Distribuidor
- US, Puerto Rico, Canada, EU, China, Brazil, Chile, and Dominican Republic.
- Lote
- REF/Product Code 1632660: Lot 3124-0108-6568, exp. 2018- 04-01; REF/Product Code 8318925: Lot 3124-0108-6569, exp. 2018- 05-01
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAVITROS Performance Verifier II3 de abril de 2024
- FDAVITROS Performance Verifier I3 de abril de 2024
- FDAVITROS XT Chemistry Products TBIL-ALKP Slides (ALKP is not affected) In Vitro Diagnostic use only - The TBIL2 de septiembre de 2020
- FDAVITROS Chemistry Product Performance Verifier I, Product Code 8231474, UDI # 10758750004577 - Product Usage:24 de junio de 2020
- FDAArtis Q and Q.Zen fluoroscopic x-ray system22 de marzo de 2017
- FDAC2 CryoBalloon Ablation System - Controller Cap. C2 CryoBalloon Controller; Cryosurgical Unit, Cryogenic22 de marzo de 2017
Preguntas sobre este retiro
¿VITROS Chemistry Products Li (Lithium) Slides, REF/Catalog Number/Product Code 1632660 (18 slide format, 90 slides per sales unit) and REF/Catalog Number/Product Code 8318925 (60 slide format / 300 slides per sales unit), Unique Identifier Nos. 10758750009466 and 10758750004720, RX ONLY, IVD --Ortho Clinical Diagnostics, Inc. Rochester, NY 14626 --- Background: The VITROS Li Slide is a multilayered, analytical element coated on a polyester support. During the coating process, thin layers of reagents are coated onto a polyester support layer to create a spreading layer and a chemical (reagent) layer. During the coating process rolls of materials are produced. These rolls are segmented into slits. There are several slits in each roll. Each slit is then cut into small chemistry chips that are positioned between plastic to create the final Microslide format. Eighteen (18) or sixty (60) slides are placed into each plastic cartridge for use. Each slide is a device. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1471-2017 el 22 de marzo de 2017. Peligro reportado: Intermittent imprecision in results generated using two specific lots of VITROS Li Slides.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter