Stryker Instruments 0400-710-000 T4 Pullover Toga (S/M) Sterile, EO; single use; Product Usage: Protection System and are intended to protect the patient, healthcare personnel and operating room personnel against contamination, exposure of infectious bodily fluids, and the transfer of microorganismsand particulate material.
Publicado el 12 de junio de 2013 · Inicio del retiro: 23 de mayo de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Based on an accelerated aging test for this product packaging, there is an indicator of potential for a breach in the sterile barrier in the header bag and potentially the blue sterilization wrap for togas within the second and third year of shelf life. The test method used involved high heat to accelerate the aging of the packaging. Stryker is conducting additional testing and assessing this pr
Qué hacer
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Detalles del retiro
- Fabricante
- Stryker Instruments Div. of Stryker Corporation
- Distribuidor
- USA Nationwide Distribution
- Lote
- 1006134, 1101020, 1106424
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States