Brilliance BigBore Oncology CT - 728243 Computed Tomography X-ray systems intended to produce cross-sectional Images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes. These devices may include signal analysts and display equipment, patient and equipment supports, components, and accessories. This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures.

Publicado el 29 de marzo de 2017 · Inicio del retiro: 23 de febrero de 2017

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

When the user performs a retrospective respiratory gated 4D CT scan with the Bellows pulmonary gating device, the CT images might be reconstructed at a single phase, while the annotations on the images incorrectly indicate that the reconstruction is from the requested specific phases (0% to 90%).

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Philips Medical Systems (Cleveland) Inc
Distribuidor
Worldwide Distribution - US to AR, CA, CO, CT, FL, GA, IL, IN, KY, LA, MA, MD, MI, MO, NC, NE, NJ, NM, NY, OH, TX & WI; Internationally to the following countries: Australia, Austria, Canada, China, Czech Republic, Germany, India, Israel, Jordan, Korea, Mexico, New Cldonia, Romania, Russian Federation, Turkey, & United Kingdom.
Lote
System Serial Number - 76002, 76008, 76010, 76011, 76016, 76020, 76021, 76022, 76023, 76024, 76025
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States