LivaNova Heater-Cooler System 3T - Product Usage: The Heater-Cooler 3T System is used to circulate water through heat exchangers to warm or cool a patient during cardiopulmonary bypass procedures lasting 6 hours or less.
Publicado el 25 de marzo de 2020 · Inicio del retiro: 24 de febrero de 2020
Resumen
FDA registra el retiro Z-1498-2020 de dispositivos médicos de Clase II, publicado el 25 de marzo de 2020 e iniciado el 24 de febrero de 2020. Empresa responsable: LivaNova USA Inc.; Estado: Finalizado; Cantidad reportada: Total Worldwide=9309 systems (2713 system in U.S.); Lote: Device Listing: E572259 Product Code Product Description Affected Serial Number range 16-02-80 Heater-Cooler 3T, 230V 16S10027 16S16874 16-02-81 Heater-Cooler 3T, 240V 16S10743 16S11708 16-02-82 Heater-Cooler 3T, 208V 16S10772 16S16840 16-02-83 Heater-Cooler 3T, 127V 16S11455 16S16509 16-02-85 Heater-Cooler 3T, 120V 16S10958 16S16847 16-02-95 Heater-Cooler 3T, 200V 16S12004 16S16818. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The firm is providing validated cleaning and disinfection instructions in their revised operating instructions to reduce the risk of NTM contamination.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- LivaNova USA Inc.
- Distribuidor
- U.S. Nationwide distribution including Puerto Rico. This matter was triggered by the clearance of K191402, thus only the United States, including Puerto Rico, is affected. O.U.S. - none
- Lote
- Device Listing: E572259 Product Code Product Description Affected Serial Number range 16-02-80 Heater-Cooler 3T, 230V 16S10027 16S16874 16-02-81 Heater-Cooler 3T, 240V 16S10743 16S11708 16-02-82 Heater-Cooler 3T, 208V 16S10772 16S16840 16-02-83 Heater-Cooler 3T, 127V 16S11455 16S16509 16-02-85 Heater-Cooler 3T, 120V 16S10958 16S16847 16-02-95 Heater-Cooler 3T, 200V 16S12004 16S16818
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDASterile Disposable Connectors ASY 1/2 CONN W/LL STRLE 24 REF/ GTIN for Insert Label:050608000/ 008036221233828 de julio de 2020
- FDAVein Irrigation Cannula with blunt tip. ASY VIC, STERILE 25/CS REF/ GTIN for Insert Label: VIC/ 008036221331078 de julio de 2020
- FDASterile Disposable Connectors ASY RDC Y 1/2X3/8X3/8 STR 24 REF/ GTIN for Insert Label:050529000/8 de julio de 2020
- FDASterile Disposable Connectors ASY STR CONN 1/4 STRLE 24. REF/ GTIN for Insert Label:050504000/ 008036221232218 de julio de 2020
- FDAArtis Q.zen Biplane, Model No. 10848355 Artis Q.zen Ceiling, Model No. 10848354 Artis Q.zen Floor, Model No.25 de marzo de 2020
- FDAAGFA Digital Radiography X- Ray system DR 800 with TomoSynthesis- a Tomographic X-ray system - Product Usage:25 de marzo de 2020
Preguntas sobre este retiro
¿LivaNova Heater-Cooler System 3T - Product Usage: The Heater-Cooler 3T System is used to circulate water through heat exchangers to warm or cool a patient during cardiopulmonary bypass procedures lasting 6 hours or less. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1498-2020 el 25 de marzo de 2020. Peligro reportado: The firm is providing validated cleaning and disinfection instructions in their revised operating instructions to reduce the risk of NTM contamination.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter