Giraffe Bedded Warmer, Panda Warmer Modifiction to Giraffe and Panda Warmer- Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SP02 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant
Publicado el 29 de marzo de 2017 · Inicio del retiro: 18 de noviembre de 2016
Resumen
FDA registra el retiro Z-1508-2017 de dispositivos médicos de Clase II, publicado el 29 de marzo de 2017 e iniciado el 18 de noviembre de 2016. Empresa responsable: Ohmeda Medical; Estado: Finalizado; Cantidad reportada: 5,993 (5,080 US; 913 OUS); Lote: See Attached Spreadsheat. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
Qué hacer
Voluntary: Firm initiated · E-Mail
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Ohmeda Medical
- Distribuidor
- Worldwide Distribution - US (nationwide and Puerto Rico) Bahamas and Internationally to UK, Trinidad & Tobago, Taiwan, Spain, Panama, Netherlands Antilles, Mexico, Lebanon, Japan, Ireland, Honduras, El Salvador, Ecuador, Costa Rica, Colombia, Canada, Brazil, and Venezuela,
- Lote
- See Attached Spreadsheat
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAGiraffe Bedded Warmer, Panda Warmer Giraffe and Panda Warmer- Infant radiant warmers provide infrared heat in29 de marzo de 2017
- FDAGiraffe Incubator Ohmeda Medical Giraffe Incubator-The Giraffe Incubator is an infant incubator. Incubators29 de marzo de 2017
- FDACarePlus Incubator Ohmeda-Ohio CarePlus Incubator- Incubators provide a controlled thermal environment for29 de marzo de 2017
- FDAGiraffe Bedded Warmer, Panda Warmer Modification to Giraffe and Panda Warmers (Giraffe and Panda29 de marzo de 2017
- FDAMammotome Revolve Dual Vacuum Assist Biopsy System The device is used to sample breast tissue that has been29 de marzo de 2017
- FDAPoly Component Trial, UCCS, packaged in the following sizes and configurations: a) Poly Comp Trial - UCCS29 de marzo de 2017
Preguntas sobre este retiro
¿Giraffe Bedded Warmer, Panda Warmer Modifiction to Giraffe and Panda Warmer- Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SpO2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SP02 sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant fue retirado del mercado?
Sí. FDA publicó el retiro Z-1508-2017 el 29 de marzo de 2017. Peligro reportado: Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where the power cord connects to the wall outlet which could cause thermal injury.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · E-Mail