Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098
Publicado el 17 de abril de 2024 · Inicio del retiro: 13 de marzo de 2024
Resumen
FDA registra el retiro Z-1518-2024 de dispositivos médicos de Clase II, publicado el 17 de abril de 2024 e iniciado el 13 de marzo de 2024. Empresa responsable: Intuitive Surgical, Inc.; Estado: En curso; Cantidad reportada: 53; Lote: Part/UDI-DI: 380601/00886874114605. SP1098 Serial Numbers: SP0202, SP0253, SP0254, SP0219, SP0228, SP0231, SP0238, SP0273, SP0208, SP0226, SP0240, SP0247, SP0249, SP0256, SP0257, SP0268, SP0270, SP0272, SP0221, SP0223, SP0225, SP0229, SP0230, SP0232, SP0233, SP0235, SP0236, SP0237, SP0239, SP0241, SP0242, SP0243, SP0244, SP0245, SP0248, SP0250, SP0251, SP0252, SP0255, SP0258, SP0259, SP0260, SP0261, SP0262, SP0263, SP0264, SP0265, SP0266, SP0269, SP0274, SP0278, SP0280, SP0282. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries, body wall injury or emergent conversion to open surgery, 2) Bottom failure: pinching injury to the user with no harm to the patient.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Intuitive Surgical, Inc.
- Distribuidor
- US: VA, CA, NH, FL, NY, CO, AL, KY, MD, NC, NJ, OH, SC, SD, TN, WA, WI, WV, MS, TX, KS. OUS: China, France, Japan, South Korea
- Lote
- Part/UDI-DI: 380601/00886874114605. SP1098 Serial Numbers: SP0202, SP0253, SP0254, SP0219, SP0228, SP0231, SP0238, SP0273, SP0208, SP0226, SP0240, SP0247, SP0249, SP0256, SP0257, SP0268, SP0270, SP0272, SP0221, SP0223, SP0225, SP0229, SP0230, SP0232, SP0233, SP0235, SP0236, SP0237, SP0239, SP0241, SP0242, SP0243, SP0244, SP0245, SP0248, SP0250, SP0251, SP0252, SP0255, SP0258, SP0259, SP0260, SP0261, SP0262, SP0263, SP0264, SP0265, SP0266, SP0269, SP0274, SP0278, SP0280, SP0282
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAda Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-1522 de julio de 2026
- FDADa Vinci ASSY, PSS, SP1098, Part Number: 3806011 de julio de 2026
- FDABrand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/A10 de junio de 2026
- FDAIntuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-0613 de mayo de 2026
- FDANEXLIZET (bempedoic acid and ezetimibe) tablets, 180 mg/10 mg, 30-count bottle, Rx only, Manufactured for:24 de abril de 2024
- FDAEndoVive 20 Fr Safety PEG Push w/ENFit Kit (Box 2)-Indicated for enteral nutrition directly into the stomach17 de abril de 2024
Preguntas sobre este retiro
¿Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098 fue retirado del mercado?
Sí. FDA publicó el retiro Z-1518-2024 el 17 de abril de 2024. Peligro reportado: Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries, body wall injury or emergent conversion to open surgery, 2) Bottom failure: pinching injury to the user with no harm to the patient.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter