Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1529-2015

MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg/BP Combo 47, Catalog B1017-417 Microbiology: For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.

Publicado el 6 de mayo de 2015 · Inicio del retiro: 8 de agosto de 2014

Resumen

FDA registra el retiro Z-1529-2015 de dispositivos médicos de Clase II, publicado el 6 de mayo de 2015 e iniciado el 8 de agosto de 2014. Empresa responsable: Beckman Coulter, Inc.; Estado: Finalizado; Cantidad reportada: Approx ~(216240 boxes U.S. and 84249 boxes OUS) total; Lote: Catalog number: B1017-417; Siemens Material Number (SMN): 10483099. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Technical Support Bulletin issued in February 2013 did not mention performance limitation for Proteus mirabilis and Imp when using the CLSIM100-S20 revised interpretive criteria.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Beckman Coulter, Inc.
Distribuidor
Worldwide Distribution: US (Nationwide) and to countries of: Argentina, Bahamas, Brazil, Canada, China, Ecuador, Egypt, Germany, Lithuania, Mexico, Paraguay, Peru Poland, Portugal, Romania, Russian Federation, Singapore, South Africa, Spain, Taiwan, Thailand, Trinidad and Tobago, United Arab Emirates, Uruguay, and Venezuela.
Lote
Catalog number: B1017-417; Siemens Material Number (SMN): 10483099.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg/BP Combo 47, Catalog B1017-417 Microbiology: For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1529-2015 el 6 de mayo de 2015. Peligro reportado: Technical Support Bulletin issued in February 2013 did not mention performance limitation for Proteus mirabilis and Imp when using the CLSIM100-S20 revised interpretive criteria.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter