Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1533-2013

BARD PERIPHERAL VASCULAR VACCESS® PTA Balloon Dilatation Catheters, 8mm x 4cm x 80cm. Product Usage: Vaccess® PTA Balloon Dilatation Catheters are recommended for use in Percutaneous Transluminal Angioplasty of the femoral, iliac and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also recommended for post-dilatation of balloon expandable stents, self-expanding stents and stent grafts in the peripheral vasculature. This catheter is not for use in coronary arteries. Vaccess® PTA Balloon Dilatation Catheters are supplied sterile and intended for single use.

Publicado el 19 de junio de 2013 · Inicio del retiro: 28 de mayo de 2013

Resumen

FDA registra el retiro Z-1533-2013 de dispositivos médicos de Clase II, publicado el 19 de junio de 2013 e iniciado el 28 de mayo de 2013. Empresa responsable: Bard Peripheral Vascular Inc; Estado: Finalizado; Cantidad reportada: 415; Lote: Product Codes and Lot Numbers, respectively: VA8084R, 93JW0056 VA8084, 937W0090. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Bard Peripheral Vascular (BPV) has confirmed that some product code/lot number combination of devices may have the incorrect balloon diameter labeled on the inflation hub. Specifically, the balloon size printed on the hub may read 7mm x 4cm, when the actual balloon size is 8mm x 4cm. All other product labeling is consistent with the actual 8mm x 4cm balloon size.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Bard Peripheral Vascular Inc
Distribuidor
US Nationwide Distribution
Lote
Product Codes and Lot Numbers, respectively: VA8084R, 93JW0056 VA8084, 937W0090
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿BARD PERIPHERAL VASCULAR VACCESS® PTA Balloon Dilatation Catheters, 8mm x 4cm x 80cm. Product Usage: Vaccess® PTA Balloon Dilatation Catheters are recommended for use in Percutaneous Transluminal Angioplasty of the femoral, iliac and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also recommended for post-dilatation of balloon expandable stents, self-expanding stents and stent grafts in the peripheral vasculature. This catheter is not for use in coronary arteries. Vaccess® PTA Balloon Dilatation Catheters are supplied sterile and intended for single use. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1533-2013 el 19 de junio de 2013. Peligro reportado: Bard Peripheral Vascular (BPV) has confirmed that some product code/lot number combination of devices may have the incorrect balloon diameter labeled on the inflation hub. Specifically, the balloon size printed on the hub may read 7mm x 4cm, when the actual balloon size is 8mm x 4cm. All other product labeling is consistent with the actual 8mm x 4cm balloon size.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter